Nihon Kohden Corporation ZM-930PA 用户手册
30
Operator's Manual ZM-920PA/930PA
Position 3
In this position, the respiration waveform is optimum, but the ECG lead is
unusual.
In this position, the respiration waveform is optimum, but the ECG lead is
unusual.
Position 4
In this position, the respiration measurement is influenced by the impedance
variation of the abdomen, so the cardiac pulse wave included in the respiration
wave is reduced. Note that the waveform is inverted in phase compared with the
chest movement (the waveform goes down during inspiration). It is difficult to
measure the ECG at the same time.
In this position, the respiration measurement is influenced by the impedance
variation of the abdomen, so the cardiac pulse wave included in the respiration
wave is reduced. Note that the waveform is inverted in phase compared with the
chest movement (the waveform goes down during inspiration). It is difficult to
measure the ECG at the same time.
Position 2
In this position, the waveform amplitude is usually large and the ECG lead is
similar to Lead MII. This position can be generally recommended.
In this position, the waveform amplitude is usually large and the ECG lead is
similar to Lead MII. This position can be generally recommended.
RA
LL
Right infraclavicular
fossa
fossa
Fifth intercostal space on
the left midaxillary line, V6
the left midaxillary line, V6
RA
LL
Right midaxillary at the
horizontal level of V4
horizontal level of V4
Fifth intercostal space on the
left midaxillary line, V6
left midaxillary line, V6
RA
LL
Lowest rib on the right
anterior axillary line
anterior axillary line
Lowest rib on the left
anterior axillary line
anterior axillary line