TempSen Electronics Company TEMPOD20B 用户手册

下载
页码 42
 
 
 
USER MANUAL 
UM-PDUB-160217 REV.D
 
 
Page 18 / 42 
Battery 
Tempod data logger uses one user replaceable Panasonic CR2450 lithium battery,   
Remaining Battery Indicator 
 
3 bars - Full 
 
2 bars - Good 
 
1 bar – Medium 
 
0 bar – Low, please replace battery NOW 
Replace Battery 
 
Please insert a coin in the groove on battery compartment cover to spin it in the direction of OPEN to open battery cover, and take 
battery out, then press “View/Mark” or “Start/stop” button several times until LCD changes to blank, otherwise volume of new battery 
would probably not be correctly recognized within one hour after battery replacement. 
 
Place a new Panasonic CR2450 battery into battery compartment with positive side upward. 
 
Put cover back to compartment in the position where triangle on cover points to the triangle on logger, then spin cover in the direction of 
CLOSE to close battery cover 
IATA Compliance 
Battery supplied with Tempod passed the Part III, SubSection 38.3 of the UN Manual of Tests and Criteria. Tempod® data logger is also 
certified to meet requirements in General Requirements and Section II of Packaging Instruction 970, IATA Dangerous Goods Regulations 58th 
Edition. Tempod® complies with RTCA-DO160G Environmental Conditions and Test Procedures for Airborne Equipment, it's allowed to used 
in air freight in active status. 
 
 
 
 
 
WARNING 
 
Please DO NOT try to recharge or heat the battery 
 
Please DO NOT use data logger (excluding external probe) in environment outside of -30.70 to avoid damage to 
battery.