8.7.1
Wired Detector Operation
Figure 7 shows the wired connection of the XRpad™ 4336 MED detector in a clinical environment.
The AC outlet shall be installed near the Interface and Power Unit XRpad™ IPU and shall be easily
accessible. The XRpad™ IPU may be mounted in an equipment enclosure. In the wired application
the XRpad™ 4336 MED detector is connected with the XRpad™ IPU which powers the
XRpad™ 4336 MED detector and is responsible for the data transfer. The XRD IPU is connected via
Cat 5e/6 with the Imaging Workstation. Make sure that the XRpad™ IPU is not connected directly
with the clinical network. The Trigger I/F of the XRpad™ IPU need to be connected with the
Generator and with the Hand switch.
The AC outlet shall be installed near the Interface and Power Unit XRpad™ IPU and shall be easily
accessible. The XRpad™ IPU may be mounted in an equipment enclosure. In the wired application
the XRpad™ 4336 MED detector is connected with the XRpad™ IPU which powers the
XRpad™ 4336 MED detector and is responsible for the data transfer. The XRD IPU is connected via
Cat 5e/6 with the Imaging Workstation. Make sure that the XRpad™ IPU is not connected directly
with the clinical network. The Trigger I/F of the XRpad™ IPU need to be connected with the
Generator and with the Hand switch.
Figure 7 Wired Connection of the XRpad™
The XRpad™ IPU communicates via a standard Gigabit Ethernet network Interface, and comes
equipped with an RJ45 interface port. Due to the overall network traffic it is recommended to use this
interface in a direct (Point-to-Point) connection with the host computer in order to achieve optimal
speed performance. The XRpad™ IPU should be connected to the host computer by one of the
PerkinElmer XRD GigE Interface Cables or a CAT5e /CAT6 (shielded twisted pair, stranded or solid
copper conductor) cable. The cable length can be up to 30m.
equipped with an RJ45 interface port. Due to the overall network traffic it is recommended to use this
interface in a direct (Point-to-Point) connection with the host computer in order to achieve optimal
speed performance. The XRpad™ IPU should be connected to the host computer by one of the
PerkinElmer XRD GigE Interface Cables or a CAT5e /CAT6 (shielded twisted pair, stranded or solid
copper conductor) cable. The cable length can be up to 30m.