8.7.1
 
Wired Detector Operation 
Figure 7 shows the wired connection of the XRpad™ 4336 MED detector in a clinical environment. 
The AC outlet shall be installed near the Interface and Power Unit XRpad™ IPU and shall be easily 
accessible.  The XRpad™ IPU may be mounted in an equipment enclosure. In the wired application 
the XRpad™ 4336 MED detector is connected with the XRpad™ IPU which powers the 
XRpad™ 4336 MED detector and is responsible for the data transfer. The XRD IPU is connected via 
Cat 5e/6 with the Imaging Workstation.  Make sure that the XRpad™ IPU is not connected directly 
with the clinical network. The Trigger I/F of the XRpad™ IPU need to be connected with the 
Generator and with the Hand switch.  
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
Figure 7  Wired Connection of the XRpad™  
 
The XRpad™ IPU communicates via a standard Gigabit Ethernet network Interface, and comes 
equipped with an RJ45 interface port. Due to the overall network traffic it is recommended to use this 
interface in a direct (Point-to-Point) connection with the host computer in order to achieve optimal 
speed performance. The XRpad™ IPU should be connected to the host computer by one of the 
PerkinElmer XRD GigE Interface Cables or a CAT5e /CAT6 (shielded twisted pair, stranded or solid 
copper conductor) cable. The cable length can be up to 30m.