8.7.3.1
Power On the XRpad™ 4336 MED Detector
This chapter describes the power up the XRpad™ 4336 MED. For more details check the Digital
Radiography Software Manual. Please make sure that the IP setting on your network adapter is set to
static IP and in correlation to the XRpad™ 4336 MED. The default settings of the XRpad™ 4336
MED are “192.168.2.158” for the LAN – connection and “192.168.22.1 for the wireless LAN –
connection. It is required that the detector has an XRpad™ Protective Insert or an XRpad™ LPB
inserted into battery compartment.
Radiography Software Manual. Please make sure that the IP setting on your network adapter is set to
static IP and in correlation to the XRpad™ 4336 MED. The default settings of the XRpad™ 4336
MED are “192.168.2.158” for the LAN – connection and “192.168.22.1 for the wireless LAN –
connection. It is required that the detector has an XRpad™ Protective Insert or an XRpad™ LPB
inserted into battery compartment.
8.7.3.2
Wired Mode
Plug in the power cord to the Interface and Power Unit XRpad™ IPU and switch the Power On. The
AC Input LED will turn on (Green) and the DC Output LED will turn ON (Yellow). To power the
XRpad™ 4336 MED detector, press the XRpad™ 4336 MED detector Power push button of the
XRpad™ IPU or the push button at the XRpad™ itself for 1s. The DC Output LED will turn from
Yellow to green.
During the initialization of the detector the detector push button LED is flashing fast. Once the
XRpad™ 4336 MED detector is powered, the XRpad™ 4336 MED detector LED will be flashing
slowly and the Detector display is showing the current status. After preparing the Radiography
Imaging Software for exposure the XRpad™ 4336 MED detector LED will turn in a continuously
ON. The XRpad™ 4336 detector LED ON will indicate that the detector is ready for exposure.
AC Input LED will turn on (Green) and the DC Output LED will turn ON (Yellow). To power the
XRpad™ 4336 MED detector, press the XRpad™ 4336 MED detector Power push button of the
XRpad™ IPU or the push button at the XRpad™ itself for 1s. The DC Output LED will turn from
Yellow to green.
During the initialization of the detector the detector push button LED is flashing fast. Once the
XRpad™ 4336 MED detector is powered, the XRpad™ 4336 MED detector LED will be flashing
slowly and the Detector display is showing the current status. After preparing the Radiography
Imaging Software for exposure the XRpad™ 4336 MED detector LED will turn in a continuously
ON. The XRpad™ 4336 detector LED ON will indicate that the detector is ready for exposure.
8.7.3.3
Wireless Mode
When the detector is not connected to the XRpad™ IPU then check the status of the XRpad™ LBP
to ensure the charge of the battery is more than 50%. If the status is low, exchange the XRpad™ LBP
with a charged one or use the wired operation mode. Press the detector push button for 1s and the
XRpad™ 4336 MED detector will be powered up. During the initialization of the detector the
detector push button LED is flashing fast. Once the detector is powered the blue LED of the push
button will be flashing slowly and the Detector display is showing the current status. After preparing
the Radiography Imaging Software for exposure the XRpad™ 4336 MED detector LED will turn in
a continuously ON. The XRpad™ 4336 MED detector LED ON will indicate that the XRpad™
4336 MED detector is ready for exposure.
to ensure the charge of the battery is more than 50%. If the status is low, exchange the XRpad™ LBP
with a charged one or use the wired operation mode. Press the detector push button for 1s and the
XRpad™ 4336 MED detector will be powered up. During the initialization of the detector the
detector push button LED is flashing fast. Once the detector is powered the blue LED of the push
button will be flashing slowly and the Detector display is showing the current status. After preparing
the Radiography Imaging Software for exposure the XRpad™ 4336 MED detector LED will turn in
a continuously ON. The XRpad™ 4336 MED detector LED ON will indicate that the XRpad™
4336 MED detector is ready for exposure.
8.7.4
Power Down the XRpad™ 4336 MED
The XRpad™ 4336 MED detector is powered OFF by holding down one of the three push buttons
for more than 4 seconds. The following buttons can be used
for more than 4 seconds. The following buttons can be used
•
XRpad™ 4336 MED detector (Figure 1 (5)) (wireless & wired mode)
•
XRpad™ IPU push button (Figure 6 (5)) (wired mode)
•
Extended hand switch push button (wired mode)