11.  Place  the  remote  unit  onto  the  patients  skin  with  the 
electrodes aligned over the stimulated zone.  
 
 
IMPORTANT: Keep the remote unit in the same position and 
direction as it was before to ensure that the measure area is 
placed directly over the previously stimulated area. 
 
 
12.  Fix  the  remote  unit  to  the  forearm  with  the  armband 
again.  
 
13.  Press  the  remote  unit  button..
 
The  “Collecting 
sweat”  phase  will  start  automatically  and  the  bips 
stop.  
 
 
The electrode position is very important. The measurement electrode must be in contact with 
the  skin,  directly  over  the  previous  stimulated  zone  (reddened  zone).  If  it  is  anywhere  else 
the  sample  of  sweat  generated  may  not  be  enough  to  give  a  result  or  the  measurement 
electrode may not have a good contact with the sample, so you won’t get a good result. 
 
Phase 3:
 Collecting sweat 
 
14.  The screen shows: 
The remote makes a repeated visual signal of two green 
flashes.  
This process lasts for 10 minutes. During this phase, the 
remote  unit  needs  to  receive  a  sufficiently  amount  of 
sample  to  be  able  to  analyze  it  (see  Technical 
specifications
 chapter).  
 
 
Phase 4:
 Measuring patient 
 
15.  Once  10  minutes  of  the  Collecting  sweat  phase  is 
over,  the  equipment  will  automatically  start  the 
Measuring  patient  phase,  where  it  will  measure  the 
chloride  concentration  of  sweat,  without  further  user 
involvement.  
 
16.  The screen shows: 
== “Measuring patient”  
 
17.  The Measuring patient process takes 5 minutes; after 
this, the equipment sends the  result from the remote 
unit to the base unit.  
 
 
CAUTION: During measuring phase, it is important that the distance between the base and 
the remote is no greater than 10 meters. If it’s any greater it could lead to a break in the link 
and the test would have to be repeated.