TECNICAS CIENTIFICAS PARA LABORATORIO S.A. T620 用户手册
ISEsweat
- User’s Manual 26
11. Place the remote unit onto the patients skin with the
electrodes aligned over the stimulated zone.
IMPORTANT: Keep the remote unit in the same position and
direction as it was before to ensure that the measure area is
placed directly over the previously stimulated area.
12. Fix the remote unit to the forearm with the armband
again.
13. Press the remote unit button..
The “Collecting
sweat” phase will start automatically and the bips
stop.
stop.
The electrode position is very important. The measurement electrode must be in contact with
the skin, directly over the previous stimulated zone (reddened zone). If it is anywhere else
the sample of sweat generated may not be enough to give a result or the measurement
electrode may not have a good contact with the sample, so you won’t get a good result.
the skin, directly over the previous stimulated zone (reddened zone). If it is anywhere else
the sample of sweat generated may not be enough to give a result or the measurement
electrode may not have a good contact with the sample, so you won’t get a good result.
Phase 3:
Collecting sweat
14. The screen shows:
The remote makes a repeated visual signal of two green
flashes.
flashes.
This process lasts for 10 minutes. During this phase, the
remote unit needs to receive a sufficiently amount of
sample to be able to analyze it (see Technical
specifications chapter).
remote unit needs to receive a sufficiently amount of
sample to be able to analyze it (see Technical
specifications chapter).
Phase 4:
Measuring patient
15. Once 10 minutes of the Collecting sweat phase is
over, the equipment will automatically start the
Measuring patient phase, where it will measure the
chloride concentration of sweat, without further user
involvement.
Measuring patient phase, where it will measure the
chloride concentration of sweat, without further user
involvement.
16. The screen shows:
== “Measuring patient”
17. The Measuring patient process takes 5 minutes; after
this, the equipment sends the result from the remote
unit to the base unit.
unit to the base unit.
CAUTION: During measuring phase, it is important that the distance between the base and
the remote is no greater than 10 meters. If it’s any greater it could lead to a break in the link
and the test would have to be repeated.
the remote is no greater than 10 meters. If it’s any greater it could lead to a break in the link
and the test would have to be repeated.