ISEsweat
 
-  User’s  Manual                                                                                 28 
 
 
 
 
4.5  Quality Control Mode (QC)-ISEtrol N and ISEtrol A   
 
As  has  already  been  said  in  this  manual,  the  usual  working  mode for  the  ISEsweat
®
  is  the 
NORMAL mode, to run sweat tests, but, to carry out Quality Control you need to change the 
Work  Mode.  To  do  this,  please  follow  the  instructions  in  the  Settings  chapter,  and  select 
Quality Control Mode (QC) ISEtrol N or ISEtrol A  as the work mode.  
 
To  verify  the  correct  operation  of  ISEsweat
®
  it  is  recommended  that  you  use  ISETROL 
controls.  The  ISETROL  controls  are  suitable  for  ISE  (electrode  ion  selective)  analysis  of 
chloride,  and  are  within  the  optimum  working  range  of  the  device  (see  Technical 
Specifications section). 
 
CAUTION:  TECIL  not  responsible  for  any  problems  arising  from  the  use  of  a  control  in  poor 
condition  or  unsuitable  for  the  technique  or  device.  Therefore,  before  using  a  control,  you 
should consult with TECIL as to its compatibility with the ISEsweat. Each hospital or laboratory 
should  decide  for  itself  the  quality  control  steps  to  be  taken  but  TECIL  recommended  to 
perform a quality control each day before the first test of your work list. 
 
 
REMEMBER:  To  avoid  confusion,  please  change  back  to  Normal  Mode  once  Quality  Control  has 
finished. 
 
Select QC Work Mode ,  ISEtrol N or ISEtrol A (see Work Mode chapter 4.1) 
 
 
The Start screen QC will appear 
 
 
 
 
 
 
 
 
1.  Take a sensor card out of the package 
 
2. 
Then  carefully  remove  the  gel  that  covers  the 
electrode on the measure side. You can remove 
it  with  bare  hands  or  with  tweezers  Clean  the 
card surface with plenty of deionized  water   and 
dry with a gauze or blotting paper. 
 
 
3.  Fit the sensor card into the remote unit connector 
slot, with the MEASURE side upwards as seen at 
the image.  
 
 
4.  Put the remote unit in a horizontal position, on an 
even  and  horizontal  surface,  with  the  electrode 
facing upwards. Wait 1 min.