TECNICAS CIENTIFICAS PARA LABORATORIO S.A. T620 用户手册
ISEsweat
- User’s Manual 29
5. Add a control solution drop directly onto the
measurement electrode
6. Press <Ent> on the base to start the QC
process. The base will emit three beeps
repetitive to confirm that the user is ready to
start. And appear the next screen
repetitive to confirm that the user is ready to
start. And appear the next screen
7. Press the remote unit button.
8. Next to each phase will be appearing OK to
indicate that the phase has been realized
9. 4 minutes later the device will calculate the
chloride value of the control
10. After obtaining the concentration value in
mmol/l, the device will automatically send the
result to the screen and the printer, if fitted.
result to the screen and the printer, if fitted.
** If Appears the screen QC not constant ( - - )
Please clean the drop and place it in another position.
Press ENT on to the base.
** If appears the screen QC not constant ( ++ )
Please ENT on to the base until the QC re-start .
NOTE: The Manufacturer should provide an acceptance range as well as the certificate
analysis for the standardized control. Take into account expiry dates.
Each laboratory will need to establish its own quality control procedures, but it is highly
recommended to run a control before the beginning of a series of analysis and each time a
new batch of sensor cards is used.
analysis for the standardized control. Take into account expiry dates.
Each laboratory will need to establish its own quality control procedures, but it is highly
recommended to run a control before the beginning of a series of analysis and each time a
new batch of sensor cards is used.
WARNING: Values that are out of range might indicate some anomalies. If after several tests,
values continue to be out of the range suggested by the manufacturer, you must contact the
manufacturer and stop using the device. Get in touch with TECIL for instructions on possibly
changing the cards, or the control, or the device.
CAUTION: TECIL cannot take responsibility for any problem arising from the use of controls
that are in bad condition, or controls that are not suitable for the device or technique. Thus,
before purchasing a control, it is highly recommended that you check its compatibility with the
device.
values continue to be out of the range suggested by the manufacturer, you must contact the
manufacturer and stop using the device. Get in touch with TECIL for instructions on possibly
changing the cards, or the control, or the device.
CAUTION: TECIL cannot take responsibility for any problem arising from the use of controls
that are in bad condition, or controls that are not suitable for the device or technique. Thus,
before purchasing a control, it is highly recommended that you check its compatibility with the
device.
Check
with
TECIL
if
you
have
any
doubts.