Innokas Yhtyma Oy VC150 用户手册
KO00065K
VC150 Vital Signs Monitor
9-19
Nellcor SpO
2
: Nellcor SpO
2
and special features
Third SpO
2
Event: Triggers SatSeconds Alarm
The SatSeconds Safety Net
The SatSeconds “Safety Net” is for patients with saturation levels frequently
falling below the limit, but not staying below the limit long enough for the Sat-
Seconds time setting to be reached. When three or more limit violations occur
within 60 seconds, an alarm sounds even if the SatSeconds time setting has not
been reached.
falling below the limit, but not staying below the limit long enough for the Sat-
Seconds time setting to be reached. When three or more limit violations occur
within 60 seconds, an alarm sounds even if the SatSeconds time setting has not
been reached.
OxiMax SPD™ Alert Parameter
WARNING
Supplemental oxygen will attenuate patterns of desaturation. A
patient’s respiratory compromise can be proportionally more severe
before patterns appear in the saturation trend. Remain vigilant when
monitoring a patient on supplemental oxygen.
patient’s respiratory compromise can be proportionally more severe
before patterns appear in the saturation trend. Remain vigilant when
monitoring a patient on supplemental oxygen.
CAUTION
Do not modify any other alarm settings while using the SPD™
parameter.
parameter.
The OxiMax SPD™ Alert (SPD) method of detecting patterns of desaturation in
adults is a function of the software within the monitoring system, which detects
repetitive occurrences of desaturation followed by resaturation. These patterns
are indicative of repetitive reductions in airflow through the upper airway and
into the lungs. With the SPD parameter enabled, the default value for
SatSeconds alarms is 100.
adults is a function of the software within the monitoring system, which detects
repetitive occurrences of desaturation followed by resaturation. These patterns
are indicative of repetitive reductions in airflow through the upper airway and
into the lungs. With the SPD parameter enabled, the default value for
SatSeconds alarms is 100.
2.5 s
25 SatSeconds
0 40 42
44
46
48
S
100
98
96
94
92
90
88
86
84
82
80
78
76
74
SpO2
10%
50
52
54