Справочник Пользователя для Innokas Yhtyma Oy VC150

Скачать
Страница из 332
KO00065K
VC150 Vital Signs Monitor
9-19
Nellcor SpO
2
Nellcor SpO
2
 and special features
Third SpO
2
 Event: Triggers SatSeconds Alarm
The SatSeconds Safety Net
The SatSeconds “Safety Net” is for patients with saturation levels frequently 
falling below the limit, but not staying below the limit long enough for the Sat-
Seconds time setting to be reached. When three or more limit violations occur 
within 60 seconds, an alarm sounds even if the SatSeconds time setting has not 
been reached.
OxiMax SPD™ Alert Parameter
WARNING
Supplemental oxygen will attenuate patterns of desaturation. A 
patient’s respiratory compromise can be proportionally more severe 
before patterns appear in the saturation trend. Remain vigilant when 
monitoring a patient on supplemental oxygen.
CAUTION
Do not modify any other alarm settings while using the SPD™ 
parameter.
The OxiMax SPD™ Alert (SPD) method of detecting patterns of desaturation in 
adults is a function of the software within the monitoring system, which detects 
repetitive occurrences of desaturation followed by resaturation. These patterns 
are indicative of repetitive reductions in airflow through the upper airway and 
into the lungs. With the SPD parameter enabled, the default value for 
SatSeconds alarms is 100.
2.5 s
25 SatSeconds
0 40  42
44
46
48
S
100
98
96
94
92
90
88
86
84
82
80
78
76
74
SpO2
10%
50
52
54