Troubleshooting
12 Troubleshooting
• If the problem cannot be resolved by the described remedial actions, stop using the
electrosurgical generator and contact Olympus.
• Be aware that repairs may only be performed by authorized service centers.
12.1 General problems
Check the following table to identify and correct failures. If the problem cannot be resolved
by the described remedial actions, contact Olympus.
Problem
Cause
Remedy
The electrosurgical
generator does not
respond when pressing
the power switch.
Improper connection of the cord set
to the socket on the rear panel of the
electrosurgical generator or to the
grounded mains electricity socket.
Check the cord set and the grounded
fixed socket-outlet for correct
connection.
The grounded fixed socket-outlet has
wrong or no output voltage.
Check the grounded fixed socket-
outlet or use an alternative grounded
fixed socket-outlet.
The cord set is damaged.
Check the cord set for damages and,
if necessary, replace the cord set.
Malfunction of the electrosurgical
generator.
Do not use this electrosurgical
generator.
Contact Olympus.
After switching on
the electrosurgical
generator the touch
screen remains dark
but sounds are audible.
Malfunction of the touch screen.
Do not use this electrosurgical
generator.
Contact Olympus.
The touch screen
cannot be controlled.
An object is in contact with the touch
screen.
Remove the object.
Malfunction of the touch screen.
Do not use this electrosurgical
generator.
Contact Olympus.
There is no sound
coming from the
electrosurgical
generator during
activation.
The volume for the output tone is set
to an inaudible level, e.g. due to high
environmental noise.
Increase the volume either on the
screen [Select Menu] or use the
volume control on the rear panel of
the electrosurgical generator.
Malfunction of the electrosurgical
generator.
Do not use this electrosurgical
generator.
Contact Olympus.
The volume cannot be
adjusted on the screen
[Select Menu] or via
the volume control
on the rear panel of
the electrosurgical
generator.
The volume of error-related alarm
tones is not adjustable.
No action required.
Malfunction of the electrosurgical
generator.
Do not use this electrosurgical
generator.
Contact Olympus.