Troubleshooting
Error code Message
Cause
Remedial action
E0214
Low battery
If the problem persists,
contact Olympus.
Malfunction of the
electrosurgical generator.
Do not use this
electrosurgical generator.
Contact Olympus.
E0438
Procedure data error
1. One or more procedures
have been deleted. Press
OK to continue.
2. If the problem persists,
contact Olympus.
One or more saved
procedures are deleted.
Tap [OK] to close the error
message and to continue.
Malfunction of the
electrosurgical generator.
Do not use this
electrosurgical generator.
Contact Olympus.
E0441
Device setting error
Device settings have been
set to default.
If the problem persists,
contact Olympus.
All settings of the
electrosurgical generator are
set to default.
The electrosurgical
generator is ready for use
after the error message is
closed.
Malfunction of the
electrosurgical generator.
Do not use this
electrosurgical generator.
Contact Olympus.
E0660
No bipolar cutting
For safety reasons the
bipolar cutting modes cannot
be activated while an RFITT
mode is selected.
A bipolar cutting mode as
well as a bipolar RFITT
mode are selected.
Switch off the bipolar RFITT
mode or select the mode
SoftCoag.
E0683
No compatible RFITT
instrument connected
RFITT modes can only be
activated if a compatible
RFITT instrument is
connected.
There is no or an
incompatible HF instrument
connected.
Connect an HF instrument
that is compatible with this
electrosurgical generator
and its RFITT modes.
E0687
Instrument wearout note
The instrument is about to
reach its maximum usage
time.
Replace the instrument if
possible.
The permitted usage span of
the HF instrument is nearly
reached. The error message
is repeated after every new
activation.
Replace the HF instrument
as soon as the treatment
situation allows.
E0688
Instrument wearout
warning
The instrument has
exceeded its maximum
usage time.
Replace the instrument to
proceed.
The permitted usage span
of the HF instrument is
exceeded. It is impossible to
activate this HF instrument
anymore.
Replace the HF instrument.
E0689
Refer to error code E0687.
E0690
Refer to error code E0688.
E0###
Error
Electrosurgical generator will
automatically restart.
If the problem persists,
contact Olympus.
Malfunction of the
electrosurgical generator.
The electrosurgical
generator automatically
restarts.
If any error occurs
permanently or repetitively,
contact Olympus.
Table: Troubleshooting – Overview of error messages