Troubleshooting
Error code Message
Cause
Remedial action
E0214
Low battery
If the problem persists, 
contact Olympus.
Malfunction of the 
electrosurgical generator.
Do not use this 
electrosurgical generator. 
Contact Olympus.
E0438
Procedure data error
1. One or more procedures 
have been deleted. Press 
OK to continue.
2. If the problem persists, 
contact Olympus.
One or more saved 
procedures are deleted.
Tap [OK] to close the error 
message and to continue.
Malfunction of the 
electrosurgical generator.
Do not use this 
electrosurgical generator. 
Contact Olympus.
E0441
Device setting error
Device settings have been 
set to default.
If the problem persists, 
contact Olympus.
All settings of the 
electrosurgical generator are 
set to default.
The electrosurgical 
generator is ready for use 
after the error message is 
closed.
Malfunction of the 
electrosurgical generator.
Do not use this 
electrosurgical generator. 
Contact Olympus.
E0660
No bipolar cutting
For safety reasons the 
bipolar cutting modes cannot 
be activated while an RFITT 
mode is selected.
A bipolar cutting mode as 
well as a bipolar RFITT 
mode are selected. 
Switch off the bipolar RFITT 
mode or select the mode 
SoftCoag.
E0683
No compatible RFITT 
instrument connected
RFITT modes can only be 
activated if a compatible 
RFITT instrument is 
connected.
There is no or an 
incompatible HF instrument 
connected.
Connect an HF instrument 
that is compatible with this 
electrosurgical generator 
and its RFITT modes.
E0687
Instrument wearout note
The instrument is about to 
reach its maximum usage 
time.
Replace the instrument if 
possible.
The permitted usage span of 
the HF instrument is nearly 
reached. The error message 
is repeated after every new 
activation.
Replace the HF instrument 
as soon as the treatment 
situation allows.
E0688
Instrument wearout 
warning
The instrument has 
exceeded its maximum 
usage time.
Replace the instrument to 
proceed.
The permitted usage span 
of the HF instrument is 
exceeded. It is impossible to 
activate this HF instrument 
anymore.
Replace the HF instrument.
E0689
Refer to error code E0687.
E0690
Refer to error code E0688.
E0###
Error
Electrosurgical generator will 
automatically restart.
If the problem persists, 
contact Olympus.
Malfunction of the 
electrosurgical generator.
The electrosurgical 
generator automatically 
restarts.
If any error occurs 
permanently or repetitively, 
contact Olympus.
Table: Troubleshooting – Overview of error messages