Clarity INT9 用户手册

下载
页码 35
Clarity - INT9 A/D converter 
 
Using the INT9 card 
15 
Channel 1 (to 4) 
For each channel of the INT9 A/D Converter 
the name of the signal can be edited in the 
Name
 
 field and the 
Set Units...
 
 button 
can be used to change other signal parametres. 
Effects of these changes can be seen in other 
fields of the respective Channel 1 (to 4) section.  
Name 
This field will be used to describe the signal 
in the 
Chromatogram
 window. 
Inversion of Signal 
Inverts the polarity of the signal from the 
detector. This function can also be used for 
easy correction of inversely connected wires. 
Bipolar 
Both positive and negative voltages will be 
measured. Unchecking the 
Bipolar
 option 
will improve the resolution of the signal by 1 
bit. On the other hand, it will also restrict 
the measuring of negative signals (i.e. below 
zero). 
Set Units... 
Invokes the 
Detector Units
 dialog. 
 
Fig. 9. 
Detector Units 
•  Quantity – Field for entering the quantity of 
the units. 
•  Units – Field for entering the unit symbol, 
along with the prefix. 
•  Coefficient – Allows to enter the coefficient 
between the selected units and mV (default). 
•  Autoprefix – Allows to enter the unit prefix. 
In graphs, units will be scaled automatically, 
while in the tables, units with the prefix 
entered will be used.