Clarity INT9 用户手册
Clarity - INT9 A/D converter
Using the INT9 card
17
Select Detector
Selects the detector whose parameters will be
Selects the detector whose parameters will be
set. If there is only one detector (channel)
configured on the Clarity Instrument, this
control will not be displayed.
Range
Selects the input voltage range (in mV). The
Range
Selects the input voltage range (in mV). The
selected range corresponds to the maximum
output voltage of the detector that the A/D
converter can still process.
Available ranges for INT9 card are
Available ranges for INT9 card are
156
,
1250
and
12 123 mV
.
Sample Rate
Depending on
Depending on
Supply Frequency
filter selected
in the
System Configuration - INT9 A/D Card Setup
dialog, the available sampling rates are
6.3
,
12.5
,
25
,
50
,
100
,
200
and
400
samples per
second (for 50Hz mains frequency filter) or
7.5
,
15
,
30
,
60
,
120
,
240
and
480
samples per
second (for 60Hz mains frequency filter).
Higher sampling rate allows for the
Higher sampling rate allows for the
measurement of narrower peaks, but it also
means a larger amount of data which will affect
the size of the resulting chromatogram and the
speed of its processing.
Sample rate sufficient for successful peak
Sample rate sufficient for successful peak
detection is about 20 samples per narrowest
peak.
A higher sampling rate also increases the
A higher sampling rate also increases the
amount of noise.
Note:
Range
and
Sample Rate
are set individually for
each detector, as opposed to
Mains Frequency
set
for the INT9 A/D card as a whole.