Rohde Schwarz Rohde & Schwarz HMO1002 2-channel oscilloscope, Digital Storage oscilloscope, 21-1002-RS00 数据表

产品代码
21-1002-RS00
下载
页码 60
4
Installation and safety instructions
1  Installation and 
safety instruc-
tions
1 .1  Symbols  
 
 
 
  (1) 
(2) 
(3)
 
 
 
  (4) 
(5) 
(6) 
(7) 
Symbol 1:   Caution, general danger zone –  
Refer to product documentation
Symbol 2:  Risk of electric shock
Symbol 3:  Ground
Symbol 4:  PE terminal
Symbol 5:  ON/OFF supply voltage
Symbol 6:  Stand by display
Symbol 7:  Ground terminal
1 .2  Setting up the instrument
As you can see from the picture, there are small feets on 
the bottom which can be folded out. Please make sure you 
have fully folded out the feet‘s in order to ensure stability 
of the instrument.
The instrument must be installed in a way, that at any time 
the disconnection of the power line is not restricted.
1 .3 Safety
The instrument fulfils the DIN EN 61010-1 (VDE 0411 part
1) regulations for electrical measuring, control and labora-
tory instruments and was manufactured and tested accor-
dingly. It left the factory in perfect safe condition. Hence it 
also corresponds to European Standard EN 61010-1 and In-
Fig. 1.1: Operating positions
ternational Standard IEC 61010-1. In order to maintain this 
condition and to ensure safe operation the user is required 
to observe the warnings and other directions for use in 
this manual. Housing, chassis as well as all measuring ter-
minals are connected to safety ground of the mains. The 
instrument is designed in compliance with the regulations 
of class I.
The exposed metal parts are checked against the mains 
poles with 2200 V
DC
 voltage. The instrument is in ac-
cordance with overvoltage category II. For safety reasons, 
the instrument may only be operated with authorized 
safety sockets. The power cord must be plugged in before 
signal circuits may be connected. Separating the grounds
is prohibited.
Never use the product if the power cable is damaged. 
Check regularly if the power cables are in perfect condi-
tion. Choose suitable protective measures and installation 
types to ensure that the power cord cannot be damaged 
and that no harm is caused by tripping hazards or from 
electric shock, for instance. If it is assumed that a safe 
operation is no longer possible, the instrument must be 
shut down and secured against any unintended operation. 
Safe operation can no longer be assumed:
 ❙in case of visible damage,
 ❙in case loose parts were noticed,
 ❙if it does not function any more,
 ❙after prolonged storage under unfavourable conditions 
(e.g. like in the open or in moist atmosphere),
 ❙after rough handling during transport (e.g. packaging that 
does not meet the minimum requirements by post office,
railway or forwarding agency).
1 .4  Intended Operation
The measuring instrument is intended only for use by per-
sonnel familiar with the potential risks of measuring electri-
cal quantities. For safety reasons, the instrument may only 
be connected to properly installed safety socket outlets. 
Separating the grounds is prohibited. The power plug must 
be inserted before signal circuits may be connected. The 
product may be operated only under the operating condi-
tions and in the positions specified by the manufacturer,
without the products ventilation being obstructed. If the 
manufacturers specifications are not observed, this can
result in electric shock, fire and/or serious personal injury,
and in some cases, death. Applicable local or national 
safety regulations and rules for the prevention of accidents 
must be observed in all work performed. 
Use the measurement instrument only with original HAMEG 
measuring equipment, measuring cables and power cord. Never 
use inadequately measured power cords. Before each measure-
ment, measuring cables must be inspected for damage and repla-
ced if necessary. Damaged or worn components can damage the 
instrument or cause injury.