Installation and safety instructions
The measuring instrument is designed for use in the fol-
lowing sectors:
 ❙Industrial sector
 ❙Home
 ❙Business and commercial sectors
 ❙Small businesses.
The measuring instrument is designed for indoor use only. 
Before each measurement, you need to verify at a known 
source if the measurement instrument functions properly.
1 .5  Ambient conditions
The allowed operating temperature ranges from +5 °C to 
+40 °C (pollution category 2). The maximum relative 
humidity (without condensation) is at 80%. During storage 
and transport, the temperature must be between –20 °C 
and +70 °C. In case of condensation during transportation 
or storage, the instrument will require approximately two 
hours to dry and reach the appropriate temperature prior 
to operation. The instrument is designed for use in a clean 
and dry indoor environment. Do not operate with high dust 
and humidity levels, if danger of explosion exists or with 
aggressive chemical agents. Any operating position may 
be used; however, adequate air circulation must be 
maintained. 
For continuous operation, a horizontal or inclined position 
(integrated stand) is preferable. The maximum operating 
altitude for the instrument is 2000 m. Specifications with
tolerance data apply after a warm up period of at least 30 
minutes at a temperature of 23 °C (tolerance ±2°C). Speci-
fications without tolerance data are average values.
1 .6  Warranty and repair
¸ instruments are subject to strict quality controls.
Prior to leaving the manufacturing site, each instrument 
undergoes a 10-hour burn-in test. This is followed by ex-
tensive functional quality testing to examine all operating 
modes and to guarantee compliance with the specified
technical data. The testing is performed with testing equip-
ment that is calibrated to national standards. The statutory 
warranty provisions shall be governed by the laws of the 
country in which the ¸ product was purchased. In case 
of any complaints, please contact your supplier.
Applicable only in EU countries:
To accelerate claims, customers in EU countries may also 
contact us directly for repairs. Our customer service is 
available for repair services even once the warranty period 
ends.
Return Material Authorization (RMA):
In any event, before returning an instrument, request a 
RMA number either via Internet (http://www.hameg.com) 
To disconnect from the mains, the rubber connector
on the back panel has to be unplugged.
Do not obstruct the ventilation holes.
or by fax. If you need technical support or a suitable origi-
nal packaging, please contact our service department:
HAMEG Instruments GmbH
Service
Industriestr. 6
D-63533 Mainhausen
Telefon: +49 (0) 6182 800 500
Telefax: +49 (0) 6182 800 501
E-Mail: service@hameg.com
Any adjustments, replacements of parts, maintenance and 
repair may be carried out only by authorized ¸ technical 
personnel. Only original parts may be used for replacing 
parts relevant to safety (e.g. power switches, power trans-
formers, fuses). A safety test must always be performed 
after parts relevant to safety have been replaced (visual in-
spection, PE conductor test, insulation resistance measure-
ment, leakage current measurement, functional test). This 
helps ensure the continued safety of the product.
1 .7 Maintenance
The display may only be cleaned with water or appropriate 
glass cleaner (not with alcohol or other cleaning agents). 
Follow this step by rubbing the display down with a dry, 
clean and lint-free cloth. Do not allow cleaning fluid to en-
ter the instrument. The use of other cleaning agents may 
damage the labeling or plastic and laquered surfaces.
1 .8  Measuring Category
This oscilloscope is designed for measurements on circuits
that are only indirectly connected to the mains or not 
connected at all. The instrument is not rated for any 
measurement category. Make sure the entry voltage of 
the analog channels CH1/CH2 does not exceed 200 V 
(peak  value),150 V
RMS
at 1 MΩ input impedance. The entry
voltage of the external trigger input (TRIG. EXT.) does not 
exceeded 100 V (peak value).The maximum value allowed 
for transient overvoltages is 200 V (peak value). To ensure 
compliance, it is necessary to only use probes that have 
been manufactured and tested in accordance with DIN EN 
61010-031in order to prevent transient overvoltages at the 
The product may only be opened by authorized and 
qualified personnel. Before any work is performed on 
the product or before the product is opened, it must be 
disconnected from the AC supply network. Otherwise, 
personnel will be exposed to the risk of an electric 
shock.
Clean the outer case of the instrument at regular intervals, using 
a soft, lint-free dust cloth. Before cleaning the instrument, please 
make sure that it has been switched off and disconnected from 
all power supplies. (e.g. mains or battery supply).
No instrument parts may be cleaned with alcohol or other 
cleaning agents.