4.  Vertical Scale 
Press Scale and use 
  to adjust the vertical scale of the operation results. 
 
5.  Center Frequency 
Press Center to adjust the frequency of frequency domain waveform relative to 
the horizontal center of screen. 
 
6.  Horizontal Scale 
Press Scale and use 
  to adjust the horizontal scale of frequency domain 
waveform. 
 
7.  Select Window Function 
Spectral leakage can be considerably decreased when a window function is used. 
DS1000Z provides four kinds of FFT window functions which have different 
characteristics and are applicable to measure different waveforms. You need to 
select the window function according to different waveforms and their 
characteristics. Press Window to select the desired window function and the 
default is “Rectangle”.   
 
Table 0-1 Window Functions 
Window 
Characteristics 
Waveforms Suitable for 
Measurement
 
Rectangle  The best frequency 
resolution; 
the poorest amplitude 
resolution;   
similar to the situation 
when no window is 
multiplied. 
Transient or short pulse, the 
signal levels before and after the 
multiplication are basically the 
same;   
Sine waveforms with the same 
amplitude and rather similar 
frequencies; 
Wide band random noise with 
relatively slowly changing 
waveform spectrum. 
Hanning 
Better frequency resolution; 
poorer amplitude 
resolution. 
Sine, periodic and narrow band 
random noise. 
Hamming 
A little bit better frequency 
resolution than Hanning 
Transient or short pulse, the 
signal levels before and after the 
multiplication are rather different.