DS1000Z User’s Guide 
6-8 
Blackman 
The best amplitude 
resolution;   
the poorest frequency 
resolution 
Single frequency signal, search 
for higher order harmonics. 
Flattop   
Measure signal preciselly.   
The signal that no accurate 
reference but require accurate 
measurements 
Triangle   
Better frequency resolution.  Narrowband signal with strong 
interference noise. 
 
8.  Set the view Mode 
Press View to select “Half” (default) or “Full” display mode.   
Harf: the source channel and the FFT operation results are displayed separately. 
The time domain and frequency domain signals are displayed clearly. 
Full Screen: the source channel and the FFT operation results are displayed in 
the same window to view the frequency spectrum more clearly and to perform 
more precise measurement.   
Note: in FFT mode and when MATH is the active channel, you can also press 
HORIZONTAL    SCALE to switch between “Half” and “Full”. 
 
9.  Set the Vertical Scale 
The unit of the vertical axis can be dB/dBm or Vrms which use logarithmic mode 
and linear mode to display vertical amplitude respectively. If you need to display 
the FFT frequency spectrum in a relatively larger dynamic range, dB/dBm is 
recommended. Press Unit to select the desired unit and the default is Vrms. 
 
10.  Press Scale Reset, the instrument will adjust the vertical scale to the best value 
according to the current configration automatically. 
 
Tips 
 
Signals with DC components or deviation would cause error or deviation of the 
FFT waveform components. To reduce the DC components, set the 
 to “AC”. 
To reduce the random noise and aliasing frequency components of repetitive or 
single pulse, set the 
 of the oscilloscope to “Average”.