MODEL 
L111 
Electrical 
Input 
One 
Input connection 
Two 4mm safety banana jacks 
Input level 
0 to 1A
AC
 
Uncertainties (50 / 60 Hz) 
 
 
Resolution 
0 to 10 mA: unspecified 
10 to 50 mA: ± (0,5% of reading + 1 mA) 
50 to 1000 mA: ± (0,5% of reading + 0,5 mA) 
0.1 mA 
Maximum input level 
1,2 A 
Input impedance 
Ω 
Sample rate 
64 samples/cycle 
Storage rate 
Programmable from 125 ms to 1 day 
Storage modes 
Start/Stop, FIFO or Extended Recording Mode* (XRM™) 
Recording length 
Programmable using DataView
®
 from 15 minutes to 8 weeks 
Memory 
240,000 measurements (512 kB). The recorded data is stored in 
non-volatile memory retained even if the battery is low or removed. 
Communication 
USB 2.0 optically isolated 
Power source ** 
2 x 1.5 V AA (LR6) cell Alkaline batteries 
Battery life 
100 hours to >45 days (dependent on storage rate/recording length) 
Mechanical 
Dimensions 
136x70x32 mm (5.38x2.75x1.28") 
Weight (with batteries) 
180 g (6.4 oz) 
Case 
UL94-V0 
Vibration 
IEC 68-2-6 (1.5 mm, 10 to 55 Hz) 
Shock 
IEC 68-2-27 (30 G) 
Drop 
IEC 68-2-32 (1 m) 
Environmental 
Operating temperature 
- 10°C to + 50°C (14°F to 122°F) 
Storage temperature 
- 20°C to + 60°C (- 4°F to + 140°F) 
Relative humidity 
up to 85% at 35°C (95°F) without condensation 
Altitude 
2000 m 
Safety 
Conformity 
EN 61010-1; 50 V CAT III; Pollution level 2 
Level of protection 
IP 40 
*See the end of chapter