Table 5. Display Descriptions (cont.)
Function Displays
Condition
Note
edw042.eps
Compensation higher
than measured value.
than measured value.
Delete compensation
or switch instrument
ON/OFF.
or switch instrument
ON/OFF.
edw043.eps
Wrong polarity on
jacks E and ES.
jacks E and ES.
Reverse polarity.
edw044.eps
Measured value
unstable.
unstable.
Unsteady noise
voltage. Try time
average measurement.
voltage. Try time
average measurement.
edw045.eps
Current in external
transformer too low.
transformer too low.
Reduce auxiliary
current spike
resistance.
current spike
resistance.
No reactions to button
control etc.
edw046.eps
Operation under faulty
conditions.
conditions.
Check batteries.
Switch ON/OFF if still
faulty, contact service.
Switch ON/OFF if still
faulty, contact service.
After
"START"
"START"
edw047.eps
Reverse orientation of
current clamp or
"upwards" current.
current clamp or
"upwards" current.
Reverse clamp.
edw048.eps
Checksum of EE
PROM incorrect.
PROM incorrect.