Fluke Fluke 4325162 用户手册
产品代码
4325162
Earth/Ground Tester
Display
11
Table 5. Display Descriptions (cont.)
Function Displays
Condition
Note
edw035.eps
Aux. current spike-
resistance is being
tested.
resistance is being
tested.
Wait for test result.
edw036.eps
Earth resistance is
being tested.
being tested.
Wait for test result.
ST
edw037.eps
Measuring circuit of
earth and
auxiliary-earth
electrode
disconnected.
earth and
auxiliary-earth
electrode
disconnected.
Check lead connection
on earth spikes
measuring lead might
be defective.
on earth spikes
measuring lead might
be defective.
ST
edw038.eps
Measuring circuit of
earth and probe
electrode
disconnected.
earth and probe
electrode
disconnected.
Check lead connection
on earth spikes
measuring lead might
be defective.
on earth spikes
measuring lead might
be defective.
edw039.eps
Maximum allowable
error exceeded
because of too high
sense or aux earth
spike resistance.
error exceeded
because of too high
sense or aux earth
spike resistance.
Try to moisten soil or
connect 2nd aux earth
spike in parallel.
connect 2nd aux earth
spike in parallel.
After
"START"
"START"
edw040.eps
Measuring range
exceeded.
exceeded.
Measured value is
higher than 300 k
higher than 300 k
Ω
.
edw041.eps
Display of measured
value exceeds LIMIT.
value exceeds LIMIT.
Measured value is
higher than set LIMIT.
higher than set LIMIT.