Chauvin Arnoux P01120073Clip-on ammeter adapter 39 mm P01120073 用户手册

产品代码
P01120073
下载
页码 72
20
Graph of typical phase shift as a function of a 50 Hz AC primary current:
see appendices, page 63.
- Rise time from 10 to 90% Vs: 
 70 µs
- Fall time from 90 to 10% Vs: 
 70 µs
- Output noise:
from DC...1 kHz
 1 mV or 1 A DC
from DC...5 kHz
 1.5 mV or 1.5 A DC
from 1 Hz...5 kHz
 500 µV rms  or 0.5 A rms
Distortion parameters
Maximum distortion of the frequency on the measurement (to be added to the error in the
reference range): 1% rdg from 65 to 440 Hz, 3.5% rdg from 440 to 2000 Hz and -3 dB from
2 kHz to 10 kHz. See graph of typical error as a function of the frequency, in appendices,
page 66.
Battery voltage: 
 0.1% / V
Temperature: 
 300 ppm /°C or 0.3% rdg /10°C
Humidity 10...85 % RH at ambient temperature: 
 0.5%
Position of a conductor of 
 20 mm : from DC...440 Hz
< 0.5%
from DC...1 kHz
< 1%
from DC...2 kHz
< 3%
from DC...2 kHz
< 10%
Adjacent conductor carrying an AC current 50 Hz, at 23 mm from the clamp: < 10 mA/A
Distortion of an external field of 400 A/m (50 Hz) on centred cable: < 1.3 A
Common mode rejection: > 65 dB A / V
Residual magnetism: < 4 mA/A
4.4/ MECHANICAL SPECIFICATIONS
- Watertightness: IP 30 in accordance with IEC 529
- Clamping capacity: 1 cable 
 39 mm (or 2 cables 
 25 mm)
1 busbar of cross section 50 x 12.5 mm
- Outside dimensions of clamp: 237 x 97 x 44 mm
- Fitted lead: 1.5 m
- Weight: 520 g
- Drop height: to IEC 68 -2-32
- Protection from shocks: 100 g in accordance with IEC 68-2-27
- Vibrations: to IEC 68-2-6
Primary current
10...200 A
200...1000 A
Phase shift
 2°
 1.5°
Phase error (45...65 Hz)