Chauvin Arnoux P01120073Clip-on ammeter adapter 39 mm P01120073 用户手册
![Chauvin Arnoux](https://files.manualsbrain.com/attachments/1c6a0da915cf6503e82f46888dd01d308d767b5d/common/fit/150/50/2eac20890e32766c458ded9a5b257340fd3c97f9e976af4f4b7655116c90/brand_logo.gif)
产品代码
P01120073
19
4.3/ METROLOGICAL SPECIFICATIONS
All the errors are indicated as a % of Vs (valur of the output voltage)
- Output impedance: 100 W
- zero adjustment: ±10 A by automatic step from 25 to 40 mA approx.
- zero adjustment: ±10 A by automatic step from 25 to 40 mA approx.
100 A range (10 mV/A)
■
Intrinsic error in the field of reference
Graphs of typical relative error: see appendices, page 57,58.
■
Phase error (45...65 Hz)
- Rise time from 10 to 90% Vs:
≤
100 µs
- Fall time from 90 to 10% Vs:
≤
100 µs
- Output noise:
from DC...1 kHz
≤
8 mV or 0.8 A DC
from DC...5 kHz
≤
12 mV or 1.2 A DC
from 0.1 Hz...5 kHz
≤
2.0 mV rms or 0.2 A rms
- Residual magnetism: < 4 mA/A
1000 A range (1 mV/A)
■
Intrinsic error in the reference range
Primary current
10...20 A
20...100 A
Phase shift
≤
3°
≤
2°
Primary current
0.5...20 A
20...100 A
100...150 A
(on DC only)
Accuracy
≤
1.5% + 5 mV
≤
1.5%
≤
2.5%
Primary current
0.5...100 A
100...1000 A
Accuracy
≤
1.5% + 1 mV
≤
2.5%
≤
4%
Graphs of typical relative error: see appendices, page 61, 62.
800 ... 1000 A AC
800 ... 1400 A DC
800 ... 1400 A DC
800 ... 1400 A peak