Texas Instruments Advanced bqMTester Multi-Station Test and Program Board Texas Instruments ADVANCED-BQMTESTER ADVANCED-BQMTESTER 数据表
产品代码
ADVANCED-BQMTESTER
Appendix E Error Code Definitions
Table 2. Error Code Definitions
Error Code
Error #
Description
Most Probable Cause
Possible Action
NO_ERROR
0
Successful (No errors)
LOST_SYNC
1
EV2x00 lost
EV2x00 has outdated
Contact TI to get EV2x00 with
synchronization
firmware or drivers are
latest firmware. Ensure latest
outdated.
drivers for EV2x00 installed.
NO_USB
2
USB Connection Missing.
No EV2x00 is connected.
Close program, reboot, and
connect EV2x00 first.
connect EV2x00 first.
BAD_PEC
3
Bad PEC on SMBus
Possible Bad hardware.
Replace EV2x00 / target board
WRONG_NUM_BYTES
5
Unexpected number of
Unexpected hardware
May need assistance from TI
bytes sent/received
behavior.
T2H_UNKNOWN
6
SMBus communication
Wrong kind of target
make sure that the target mode
terminated unexpectedly /
connected or target timing is
accepts the SMB command
timed out or the bus was
off Trim oscillator
being sent
busy.
INCORRECT_PARAM
7
Invalid parameter type
Incorrect parameter in call to
Contact TI
passed to function -
function. Software Bug or
especially Variant
overflow
argument.
TIMEOUT_ERROR
8
USB Timeout
No response on USB
EV2x00 or driver problems or
software is not supposed to
wait for a response.
software is not supposed to
wait for a response.
INVALID_DATA
9
AssemblePacket could not
Bad data / bad packet.
Possible version incompatibility
build a valid packet
Software found problem with
between BqTester and Module
data
under test.
ERR_UNSOLICITED_PKT
10
Found an unsolicited
Unexpected packet received.
Make corrections to software
non-error packet when
The packet may be a
looking for error packets
response from a previous
transaction that failed or that
did not check the response.
transaction that failed or that
did not check the response.
COMPARE_DIFFERENT
11
Comparison failed and
Flash comparison results in
Module under test Flash failure
data read is different from
mismatch. Possible Flash
srec
failure or SMBus failure.
BQ80XRW_OCX_INTERNAL_ERROR
12
Problems with pointers
Possible software bug or
Contact TI
being NULL etc.
overflow.
USER_CANCELLED_OPERATION
34
User clicked on cancel
button on progress bar
dialog
button on progress bar
dialog
DF_CHECKSUM_MISMATCH
51
Data Flash checksum
Flash comparison results in
Module under test Flash failure
mismatch
mismatch. Possible Flash
failure or SMBus failure.
failure or SMBus failure.
IF_CHECKSUM_MISMATCH
52
Instruction Flash
Flash comparison results in
Module under test Flash failure
checksum mismatch
mismatch. Possible Flash
failure or SMBus failure.
failure or SMBus failure.
OPERATION_UNSUPPORTED
53
Unsupported type
Software problem
Check that Module under test
and bqTester versions are
compatible. Then contact TI
and bqTester versions are
compatible. Then contact TI
ERR_TOO_MANY_QUERIES
81
Not used
ERR_BAD_QUERY_ID
82
Not used
BAD_CRC
83
Packet was corrupted
Too much noise or bad
during USB
connection
communication
ERR_TOO_MANY_RESPONSES
84
Not used
ERR_NO_QUERIES_TO_DELETE
85
Not used
ERR_QUERY_UNAVAILABLE
86
Not used
ERR_NO_RESPONSES_TO_DELETE
87
Not used
ERR_RESPONSE_UNAVAILABLE
88
Not used
ERR_TMMT_NO_RESPONSE
90
Not used
T2H_ERR_TIMEOUT
92
SMBus communication
Wrong kind of target
make sure that the target mode
terminated unexpectedly /
connected or target timing is
accepts the SMB command
timed out or the bus was
off Trim oscillator
being sent
busy.
37
SLUU397A
–
February 2011
–
Revised May 2011
Error Code Definitions
Copyright
©
2011, Texas Instruments Incorporated