Texas Instruments Advanced bqMTester Multi-Station Test and Program Board Texas Instruments ADVANCED-BQMTESTER ADVANCED-BQMTESTER 데이터 시트

제품 코드
ADVANCED-BQMTESTER
다운로드
페이지 62
Appendix E Error Code Definitions
Table 2. Error Code Definitions
Error Code
Error #
Description
Most Probable Cause
Possible Action
NO_ERROR
0
Successful (No errors)
LOST_SYNC
1
EV2x00 lost
EV2x00 has outdated
Contact TI to get EV2x00 with
synchronization
firmware or drivers are
latest firmware. Ensure latest
outdated.
drivers for EV2x00 installed.
NO_USB
2
USB Connection Missing.
No EV2x00 is connected.
Close program, reboot, and
connect EV2x00 first.
BAD_PEC
3
Bad PEC on SMBus
Possible Bad hardware.
Replace EV2x00 / target board
WRONG_NUM_BYTES
5
Unexpected number of
Unexpected hardware
May need assistance from TI
bytes sent/received
behavior.
T2H_UNKNOWN
6
SMBus communication
Wrong kind of target
make sure that the target mode
terminated unexpectedly /
connected or target timing is
accepts the SMB command
timed out or the bus was
off Trim oscillator
being sent
busy.
INCORRECT_PARAM
7
Invalid parameter type
Incorrect parameter in call to
Contact TI
passed to function -
function. Software Bug or
especially Variant
overflow
argument.
TIMEOUT_ERROR
8
USB Timeout
No response on USB
EV2x00 or driver problems or
software is not supposed to
wait for a response.
INVALID_DATA
9
AssemblePacket could not
Bad data / bad packet.
Possible version incompatibility
build a valid packet
Software found problem with
between BqTester and Module
data
under test.
ERR_UNSOLICITED_PKT
10
Found an unsolicited
Unexpected packet received.
Make corrections to software
non-error packet when
The packet may be a
looking for error packets
response from a previous
transaction that failed or that
did not check the response.
COMPARE_DIFFERENT
11
Comparison failed and
Flash comparison results in
Module under test Flash failure
data read is different from
mismatch. Possible Flash
srec
failure or SMBus failure.
BQ80XRW_OCX_INTERNAL_ERROR
12
Problems with pointers
Possible software bug or
Contact TI
being NULL etc.
overflow.
USER_CANCELLED_OPERATION
34
User clicked on cancel
button on progress bar
dialog
DF_CHECKSUM_MISMATCH
51
Data Flash checksum
Flash comparison results in
Module under test Flash failure
mismatch
mismatch. Possible Flash
failure or SMBus failure.
IF_CHECKSUM_MISMATCH
52
Instruction Flash
Flash comparison results in
Module under test Flash failure
checksum mismatch
mismatch. Possible Flash
failure or SMBus failure.
OPERATION_UNSUPPORTED
53
Unsupported type
Software problem
Check that Module under test
and bqTester versions are
compatible. Then contact TI
ERR_TOO_MANY_QUERIES
81
Not used
ERR_BAD_QUERY_ID
82
Not used
BAD_CRC
83
Packet was corrupted
Too much noise or bad
during USB
connection
communication
ERR_TOO_MANY_RESPONSES
84
Not used
ERR_NO_QUERIES_TO_DELETE
85
Not used
ERR_QUERY_UNAVAILABLE
86
Not used
ERR_NO_RESPONSES_TO_DELETE
87
Not used
ERR_RESPONSE_UNAVAILABLE
88
Not used
ERR_TMMT_NO_RESPONSE
90
Not used
T2H_ERR_TIMEOUT
92
SMBus communication
Wrong kind of target
make sure that the target mode
terminated unexpectedly /
connected or target timing is
accepts the SMB command
timed out or the bus was
off Trim oscillator
being sent
busy.
37
SLUU397A
February 2011
Revised May 2011
Error Code Definitions
Copyright
©
2011, Texas Instruments Incorporated