Analogue reading: to get the resistance in Ω, multiply the reading
on the 50 scale by the reading coefficient shown in the table.
on the 50 scale by the reading coefficient shown in the table.
7 - RESISTANCES
Never test a resistance on a live circuit.
If the “Voltest” light comes on, presence of a voltage U > 10 V
AC/DC. Unplug the leads immediately and cut off the voltage
before making the resistance measurement.
AC/DC. Unplug the leads immediately and cut off the voltage
before making the resistance measurement.
■
Connect to the terminals of the circuit or the component to test.
Do not use the A AC input on unprotected current transformers
From 7 A to 10 A, limited to 10 minutes on, 5 minutes off up to
40 °C maximum
40 °C maximum
(1) At the terminals (2) As % of the reading (rdg) from 5 to 100 % of the range
(3) As % of end of scale
(3) As % of end of scale
A DC
400 µA
4 mA
40 mA 400 mA
1 A
10 A
Voltage drop
(1)
400 mV
450 mV 650 mV
1 V
Digital
Max reading
399.9 µA 3.999 mA 39.99 mA 399.9 mA
1.000 A
10.00 A
Accuracy
(2)
1.5 % rdg ± 2 counts
Max reading
500 µA
5 mA
50 mA
500 mA
1 A
10 A
Scale
50
10
Analogue
Reading
coefficient
coefficient
x 10
x 0.1
x 1
x 10
x 0.1
x 1
Accuracy
(3)
2.5 %
Protection
Fuse 1 A HPC
Fuse 10 A
HBC
NB: Accuracies given in the frequency reference range 35 to 500 Hz.
(1) and (2) Additional error of 1 % for a crest factor FC ≤ 3 (3) At the terminals
(4) As % of the reading (rdg) from 5 to 100 % of the range (5) As % of end of scale
(1) and (2) Additional error of 1 % for a crest factor FC ≤ 3 (3) At the terminals
(4) As % of the reading (rdg) from 5 to 100 % of the range (5) As % of end of scale
V AC
and
V AC/DC
400 µA
4 mA
40 mA 400 mA
1 A
10 A
Voltage drop
(3)
400 mV
450 mV 650 mV
1 V
Max reading
399.9 µA 3.999 mA 39.99 mA 399.9 mA
1.000 A
10.00 A
Accuracy
2 % rdg ± 5 counts
Digital
on AC
(4)
Accuracy
4.5 % rdg ± 5 counts
on AC/DC
(4)
Max reading
500 µA
5 mA
50 mA
500 mA
1 A
10 A
Scale
50
10
Analogue
Reading
coefficient
coefficient
x 10
x 0.1
x 1
x 10
x 0.1
x 1
Accuracy
(5)
2.5 %
Protection
Fuse 1 A HPC
Fuse 10 A
HBC
(1) (2)
■
Digital reading: if necessary, press the following keys,
HOLD to hold in memory and for display backlighting
From 7 A to 10 A, limited to 10 minutes on, 5 minutes off up to
40 °C maximum
40 °C maximum