Lecroy WAVEACE 2034 4-channel oscilloscope, Digital Storage oscilloscope, WaveAce 2034 用户手册
产品代码
WaveAce 2034
WaveAce Remote Control
52
920836 Rev A
SAVE/RECALL SETUP - *SAV
Description
The *SAV command stores the current state of your instrument in non-volatile internal
memory. The *SAV command stores the complete front panel setup of the oscilloscope at the
time the command is issued.
memory. The *SAV command stores the complete front panel setup of the oscilloscope at the
time the command is issued.
NOTE: Communication parameters (those modified by the commands COMM_HEADER
and WAVEFORM_SETUP) are not saved when *SAV is used.
and WAVEFORM_SETUP) are not saved when *SAV is used.
Command Syntax
*SAV <panel_setup>
<panel_setup>:= 1 to 6
Example
The following instruction saves the current instrument setup in panel setup 3:
*SAV 3
Related Commands
PANEL_SETUP
SAVE/RECALL SETUP - RECALL_PANEL, RCPN
Description
The RECALL_PANEL command recalls a front panel setup from the root directory on an external
storage device (USB drive).
storage device (USB drive).
Command Syntax
ReCall_PaNel DISK,UDSK,FILE,'<filename>'
<filename>:=
A string of up to characters with the extension .SET.
Example
The following instruction recalls the front panel setup from file P012.SET on the USB drive:
RCPN DISK,UDSK,FILE,'P012.SET'
Related Commands
PANEL_SETUP, *SAV