Chapter 8. Servicing the PT878
Transport® Model PT878 Portable Liquid Flowmeter User’s Manual
163
8.4.3 Displaying the Receive Signal in Graphical Format
Used chiefly for diagnostic purposes, the Graph option helps to determine 
why the thickness gauge is not working if you suspect a problem. The graph 
shows an image of the acoustic signal. If the display does not show a signal 
image similar to Figure 110 below, you may have a problem with the 
transducer, couplant, or the programmed values. Also, some materials (such 
as PTFE
 
or fiberglass) do not support acoustic signals.
To open the Graph window and display the receive signal graphically, press 
the [
W
] or [
X
] arrow key until you reach the Graph tab and press 
[ENTER]
Figure 110: Thickness Gauge Graph Window
Press 
[F2]
 (Cancel) or 
[F3]
 (OK) to return to Operate Mode, or press the [
X
arrow key to move to another window in the option. However, if you wish 
to adjust the graph, press the [
T
] key. The screen now appears similar to 
Display
Thickness Gauge Measure
Cancel
OK
Velocity
Graph
Zero
Signal
Material