Chapter 8. Servicing the PT878
Transport® Model PT878 Portable Liquid Flowmeter User’s Manual
193
8.10.7 Testing the Battery
The Battery Test option provides additional battery information for service 
troubleshooting. To enter the option, scroll down to the Battery Test entry in 
the Service Menu and press 
[ENTER]
. The screen appears similar to 
Figure 128: Battery Test Option
The screen displays current status (Fast Charge, On Charge, Discharging, or 
On Battery), the time remaining for the PT878 to run on the battery, the time 
for the Fast Charge, and the current condition of the backup battery.
1.
If you wish to open the Part Number window, press 
[ENTER]
. You can 
choose from a 1.8 Ahr NiCd battery (part number 200-058) or a 3.0 Ahr 
NiMH battery (part number 200-081). Scroll to the battery type you 
have installed and press 
[ENTER]
.
2.
To condition the batteries, press the [ENTER] key. The “Condition 
Battery” window should now read “Stop.” The status line should 
change to “Discharging.” (Updating the status could take up to 30 sec.)
3.
To stop the discharge cycle, press the [ENTER] key and the “Condition 
Battery” window reappears. The status now changes to “On Charger.”
4.
Press 
[F3]
 (OK) to return to Operate Mode.
Service Battery Form
Cancel
OK
Status: Fast Charge
Batt Voltage     5.78 Volts
Fast Charge     3 Min
Condition Battery
Run Time          250 Minutes 
Backup Battery: OK Batt temp 27.7 °C
       Part Number:
1.8Ahr NiCd (200-058)