Chapter 3. Programming Site Data
Transport® Model PT878 Portable Liquid Flowmeter User’s Manual
23
3.2 Entering Transducer Parameters
To enter the Transducer option, scroll to the Transducer entry on the 
Program Menu and press 
[ENTER]
. The screen appears similar to Figure 10 
below. To step through each parameter, press the [
T
] key. Refer to 
Note:
Refer to the Liquid Transducer Installation Guide (916-055) for 
additional information about transducers and configurations.
.
Figure 10: Transducer Option Window
1.
The first prompt asks you to select whether you are using a wetted or a 
clamp-on transducer.
a.
Use the [
W
] and [
X
] keys to scroll between the two types.
b.
Press 
[ENTER]
 to confirm the choice.
Transducer
Pipe
Lining
Path
Transducer/Pipe
OK
Cancel
Fluid
Type:         Wetted
Clamp-on
Transducer  Special
Frequency  2.00
Tw
14
MHz
Wedge Ang 
50
°
Wedge Tmp 
25
°C
Wedge SS
1219.2
m/s
µs