Clinical setup menu parameters
*
Parameter
Modes
Range
Description
CP
AP
AutoSet
VAuto
S
ST
T
PA
C
iV
APS
ASV
ASV
Auto
Settings
Circuit
Mask type
9 9 9 9 9 9 9 9 9 9 Full Face / Nasal 
/ Pillows / Nasal 
Ultra / Pediatric
Default: Full Face
Selects the type of mask used 
by the patient.
Tube type
9 9 9 9 9 9 9 9 9 9 SlimLine / 
Standard / 3 m
Default: Standard
Shows the type of air tubing 
used by the patient. 
Not displayed if ClimateLine or 
ClimateLine
MAX
 is connected
AB filter
9 9 9 9 9 9 9 9 9 9 No  / Yes
Default: No
Enables or disables antibacterial 
filter.
Not displayed if H5i is 
connected.
Ext. 
humidifier
9 9 9 9 9 9 9 9 9 9 No  / Yes
Default: No
Enables or disables an external 
humidifier.
Therapy
Mode
9 9 9 9 9 9 9 9 9 9 CPAP / S / ST / 
T / PAC / iVAPS / 
ASVAuto / ASV / 
AustoSet / VAuto
Default: CPAP
Sets the therapy mode available 
on the device.
Set pressure 9
4–20 cm H
2
O
Default:  
4 cm H
2
O
Sets the fixed treatment 
pressure.
EPR Level
9 9
Off / 1 / 2 / 3 
Default: 3
Sets the EPR value.
Height 
9
44-100 inches 
(110–250 cm)
Default: 70 inches 
(175 cm)
Body height needed for the 
dead space determination.
Target Pt 
Rate
9
8–30 BPM
Default:
 15 BPM
The rate input to the iVAPS 
algorithm. This should be set at 
the patient’s actual respiratory 
rate.
Target Va  
9
1–30 L/min
Default: 5.2 L/min
Used to determine the amount 
of pressure support required by 
the iVAPS algorithm.
IPAP
9 9 9 9 
4–30 cm H
2
O
Default:  
10 cm H
2
O
Sets the pressure which will be 
delivered to the patient when 
the device is triggered into 
inspiration.
*Default settings may differ from country to country