Parameter
Modes
Range
Description
CP
AP
AutoSet
VAuto
S
ST
T
PA
C
iV
APS
ASV
ASV
Auto
EPAP
9 9 9 9 9 9
3–25 cm H
2
O**
Default:  
4 cm H
2
O
Sets the pressure which will 
be delivered to the patient 
when the device is cycled into 
expiration.
Dependent on IPAP.
Min PS
9 9 9 0–20 cm H
2
O
Default:  
4 cm H
2
O
Minimum pressure support in 
iVAPS mode.
Dependent on EPAP.
Max PS
9 9 9 0–28 cm H
2
O
Default:  
20 cm H
2
O
Maximum pressure support in 
iVAPS mode.
Dependent on EPAP and Min 
PS.
Respiratory 
Rate
9 9 9
5–50 BPM
Default: 10 BPM
Sets the breaths per minute 
(BPM) or ‘backup’ rate.
Ti Max
9 9 9
9
0.3–4.0 sec
Default: 2 sec
Sets the maximum limit on the 
time the device spends in IPAP.
Dependent on Respiratory Rate.
Ti Min
9 9 9
9
0.1–4.0 sec
Default: 0.3 sec
Sets the minimum limit on the 
time the device spends in IPAP.
Dependent on Ti Max.
Ti
9 9
0.1–4.0 sec
Default: 2 sec
Sets the duration of inspiration 
in timed breath.
Dependent on Respiratory Rate.
Rise Time
9 9 9 9 9
Min /  
100–900 ms
Default: 150 ms
Sets the time taken for the 
device to reach to IPAP.
Dependent on Ti Max and Ti.
Trigger
9 9 9
9 9
Very Low / Low / 
Med / High / Very 
High
Default: Med
Sets the level of inspiratory flow 
above which the device changes 
from EPAP to IPAP.
Cycle
9 9 9
9
Very Low / Low / 
Med / High / Very 
High
Default: Med
Sets the level of inspiratory flow 
below which the device changes 
from IPAP to EPAP.
Max IPAP
9
4–25 cm H
2
O
Default:  
25 cm H
2
O
Sets the maximum inspiratory 
pressure delivered by the 
device.
Min EPAP
9
4–[Max IPAP] 
cm H
2
O
Default:  
4 cm H
2
O
Sets the minimum expiratory 
pressure delivered by the 
device.
Pressure 
Support
9
0–10 cm H
2
O
Default:  
4 cm H
2
O
Difference between IPAP and 
EPAP. Adjust for patient comfort.
** 2-25 cm H
2
O for Europe and APAC