Teledyne Model Vehicle T400 用户手册

下载
页码 396
Getting Started 
Teledyne API – Model T400 Photometric Ozone Analyzer 
40 
 
Table 3-3:  Rear Panel Description 
Component Function 
cooling fan 
Pulls ambient air into chassis through side vents and exhausts through rear.  
 
AC power 
connector 
Connector for three-prong cord to apply AC power to the analyzer.  
CAUTION! The cord’s power specifications (specs) MUST comply with the power 
specs on the analyzer’s rear panel Model number label  
Model/specs label 
Identifies the analyzer model number and provides power specs 
SAMPLE 
Connect a gas line from the source of sample gas here.   
Calibration gases are also inlet here on units with the zero/span valve option installed. 
EXHAUST 
Connect an exhaust gas line of not more than 10 meters long here that leads outside 
the shelter or immediate area surrounding the instrument. 
SPAN  
On units with zero/span valve option installed, connect a gas line to the source of 
calibrated span gas here. 
ZERO AIR 
Internal Zero Air: On units with zero/span valve option installed connect the source of 
zero air here. 
DRY AIR 
On units with zero/span valve option installed connect the source of dry air here  
(- <20
C dew point). 
RX TX 
LEDs indicate receive (RX) and transmit (TX) activity on the when blinking. 
COM 2 
Serial communications port for RS-232 or RS-485.  
RS-232 
Serial communications port for RS-232 only 
DCE DTE 
Switch to select either data terminal equipment or data communication equipment 
during RS-232 communication.  
STATUS 
For ouputs to devices such as Programmable Logic Controllers (PLCs).  
ANALOG OUT 
For voltage or current loop outputs to a strip chart recorder and/or a data logger.  
CONTROL IN 
For remotely activating the zero and span calibration modes.  
ALARM 
Option for concentration alarms and system warnings.  
ETHERNET 
Connector for network or Internet remote communication, using Ethernet cable. 
ANALOG IN 
Option for external voltage signals from other instrumentation and for logging these 
signals. 
USB 
Connector for direct connection to laptop computer, using USB cable.  
Information Label 
Includes voltage and frequency specifications 
 
 
 
 
06870C DCN6332