3.3. CONNECTIONS AND SETUP 
This section presents the electrical (Section 3.3.1) and pneumatic (Section 3.3.2) 
connections for setup and preparing for instrument operation.  
3.3.1. ELECTRICAL CONNECTIONS 
Note 
To maintain compliance with EMC standards, it is required that the cable 
length be no greater than 3 meters for all I/O connections, which include 
Analog In, Analog Out, Status Out, Control In, Ethernet/LAN, USB, RS-232, 
and RS-485. 
This section provides instructions for basic connections and for options. 
3.3.1.1. Connecting Power 
Attach the power cord to the analyzer and plug it into a power outlet capable of carrying 
at least 10 A current at your AC voltage and that it is equipped with a functioning earth 
ground. 
 
 
WARNING - ELECTRICAL SHOCK HAZARD 
HIGH VOLTAGES ARE PRESENT INSIDE THE ANALYZERS CASE 
 
POWER CONNECTION MUST HAVE FUNCTIONING GROUND 
CONNECTION. 
 
DO NOT DEFEAT THE GROUND WIRE ON POWER PLUG. 
 
TURN OFF ANALYZER POWER BEFORE DISCONNECTING OR  
CONNECTING ELECTRICAL SUBASSEMBLIES. 
 
DO NOT OPERATE WITH COVER OFF. 
 
 
CAUTION - GENERAL SAFETY HAZARD 
THE T400 ANALYZER CAN BE CONFIGURED FOR BOTH 100-130 V AND  
210-240 V AT EITHER 50 OR 60 HZ.   
TO AVOID DAMAGE TO YOUR ANALYZER, MAKE SURE THAT THE AC POWER 
VOLTAGE MATCHES THE VOLTAGE INDICATED ON THE ANALYZER’S REAR 
PANEL MODEL/SERIAL NUMBER/VOLTAGE SPECS LABEL BEFORE PLUGGING 
THE T400 INTO LINE POWER.
 
 