6.7.5. Auto Range Mode (AUTO) 
In AUTO range mode, the analyzer automatically switches the reporting range between two user-
defined ranges (low and high). The unit will switch from low range to high range when the CO
2
 
concentration exceeds 98% of the low range span. The unit will return from high range back to 
low range once both the CO
2
 concentration falls below 75% of the low range span. 
In AUTO Range mode the instrument reports the same data in the same range on both the A1 
and A2 outputs and automatically switches both outputs between ranges as described above. 
Also, the RANGE test function displayed on the front panel will automatically switch to show 
which range is in effect. 
The high/low range status is also reported through the external, digital status bits (Section 
6.13.1.1).  
To set individual ranges press the following keystroke sequence. 
 
 
 
 
 
SAMPLE*         RANGE = 500.000 PPM        CO2 =X.XXX 
 
 < TST  TST >  CAL  
 
 
SETUP 
 
 
 
SETUP X.X 
 
      
PRIMARY SETUP MENU
 
  
 
 CFG  DAS
  RNGE  
PASS  CLK  MORE  
EXIT
 
 
 
 
SETUP X.X 
RANGE CONTROL MENU 
 
 
 
 
MODE
 
 
 
 
 
 
SET  UNIT
 
 
 
 
EXIT
 
 
 
 
SETUP X.X 
RANGE MODE: 
SNGL
 
 
 
 
 SNGL  DUAL 
 AUTO
 
 
     ENTR  EXIT
 
SAMPLE          
ENTER SETUP PASS : 818
 
 
        
 8        1        8 
 
                      
ENTR
  
 EXIT
 
EXIT x 2 returns
to the main 
SAMPLE display
 
 
 
SETUP X.X 
RANGE MODE: 
AUTO
 
 
 
 
 SNGL
 
 
 
 
 
 
DUAL  AUTO 
 
     ENTR
  EXIT
 
 
 
 
SETUP X.X 
RANGE CONTROL MENU 
 
 
 
 MODE
 
 
 
 
 
 
SET
  UNIT
 
 
 
 
EXIT
 
 
 
 
SETUP X.X 
HIGH RANGE: 
500.0 Conc
 
 
 
  0          0         5          0          0          .0
       
ENTR
 
 EXIT 
 
 
 
SETUP X.X 
LOW RANGE: 
500.0 Conc
 
 
 
 
 
0          0         5          0          0          .0 
     
ENTR
 
 EXIT 
Toggle the numeral 
keys to set the 
LOW and HIGH 
range value. 
ENTR accepts the 
new setting, EXIT 
ignores the new 
setting. 
 
CAUTION 
In AUTO range mode the LOW and HIGH ranges have separate slopes and offsets for 
computing CO2 concentration. 
The two ranges must be independently calibrated. 
 
NOTE 
Avoid accidentally setting the low range of the instrument with a higher span limit than 
the high range.  This will cause the unit to stay in the low reporting range perpetually 
and defeat the function of the AUTO range mode.