Model GFC7000E Instruction Manual 
Operating Instructions 
04584 Rev A1 
62 
 
6.7.6. Range Units 
The MGFC7000E can display concentrations in parts per billion (10
9
 mols per mol, PPB), parts per 
million (10
6
 mols per mol, PPM), micrograms per cubic meter (µg/m
3
UG), milligrams per cubic 
meter (mg/m
3
MG) or percent (volume CO
2
/volume sample gas, %). Changing units affects all of 
the display, analog outputs, COM port and iDAS values for all reporting ranges regardless of the 
analyzer’s range mode.  
NOTE 
Concentrations displayed in mg/m3 and ug/m3 use 0°C, 760 mmHg for Standard 
Temperature and Pressure (STP).    Consult your local regulations for the STP used by 
your agency.   
Conversion factors from volumetric to mass units are: 
CO
2
: ppb x 1.96 = µg/m3; ppm x 1.96 = mg/m3 
To change the concentration units: 
 
 
 
 
SETUP X.X       
CONC UNITS: %
 
PPM   PPB
   
UGM
   
MGM    %
 
 
    ENTER   EXIT
SAMPLE              RANGE = 500.00 PPB        CO2=X.XXX 
 
 < TST  TST >  CAL  
 
 
SETUP
 
 
 
SETUP X.X 
 
      
PRIMARY SETUP MENU
 
  
 
 CFG  DAS
  RNGE  
PASS  CLK  MORE  
EXIT
 
 
 
 
SETUP X.X 
RANGE CONTROL MENU
 
 
 
 
 
MODE
 
 
 
 
 
 
SET  
UNIT 
 
 
 
EXIT
 
 
 
 
SETUP X.X          
CONC UNITS: PPM
 
PPM   PPB
   
UGM
   
MGM    
 
   
 ENTER
   EXIT
SAMPLE          
ENTER SETUP PASS : 818
 
 
        
 8        1        8 
 
                      
ENTR
  
 EXIT
 
Select the preferred 
concentration unit. 
EXIT returns 
to the main menu. 
ENTR accepts 
the new unit, 
EXIT returns 
to the SETUP 
menu. 
 
NOTE 
Once the units of measurement have been changed the unit MUST be recalibrated, as 
the “expected span values” previously in effect will no longer be valid.  Simply entering 
new expected span values without running the entire calibration routine is not 
sufficient.