26
Standard Operations
2. Select Long to change the number of audio signals. 
Use the alphanumeric keypad to enter/specify the number of 
audio signals (4–25).
Note: A high number of audio signals increases the duration of the 
tones and a small number of audio signals reduces the duration of the 
tones.
3. Press OK to confirm. The selected number of the audio signals 
sounds as a corresponding acoustic signal.
4. Press OK to confirm. The instrument will return to 
Instrument Setup.
5.2.5 Lamp Control
The tungsten lamp produces light in the visible spectrum 320 to 
1100 nm. 
The deuterium lamp (UV-lamp) available produces light in the 
ultraviolet spectrum 190 to 360 nm.
In the overlap zone from 320 to 360 nm, either the deuterium lamp 
(UV-lamp) or the tungsten lamp can be used for measurements.
The lifetime of the lamps is influenced 
by on-off operation and the length of time in use.
Typical use is to turn the instrument on for the entire 8–10 hour 
shift, and then off until the next day.
Note: In general, avoid on/off cycles of the lamp; this shortens the lamp’s 
life span. For maximum life, turn the lamp off only if it will remain off for at 
least 4–5 hours.
The lamp switches on automatically
if a lamp is needed for the selected program
if the instrument is operating inside the lamp’s spectrum
5.2.5.1 VIS-Lamp and UV-Lamp
1. Press Lamp Control in the Instrument Setup.
2. Select On to switch on the VIS-Lamp and/or UV-Lamp.
3. Select Save: in order to define a time interval for the burning 
time of the VIS lamp and/or UV lamp.
4. Press the field below Save to select the lamp burning time.