Standard Operations
5. Select the length of time the lamp will be switched on.
Note: After this period of time the lamp automatically turns off, after no 
measurement has been made outside the corresponding range (UV/VIS). 
6. Touch OK to confirm.
7. Touch UV Switch key to select the wavelength value between 
320 nm and 360 nm, at which the instrument changes from the 
visible to UV source.
8. Use the alphanumeric keypad to enter the maximum 
wavelength for UV operation.
9. Touch OK to confirm and return to “Lamp Control“.
10. Touch OK to confirm and return to “Instrument Setup“.
Note: The selected program that requires a lamp has highest priority. If the 
lamp in the Lamp Control was turned off, the lamp will be turned on 
automatically if it is needed for the stored program.
5.2.5.2 Effect caused by the lamp settings on the Measuring Mode
UV-Lamp switches on manually. The icon ”UV Lamp“ appears in the 
display and flashes. 
The instrument will display ”Lamp Warmup...“. This takes about 
3 minutes.