Standard Operations
5.5.6.2 Update an existing barcode test
Note: Only in very few cases does the revision of a test require all test 
specifications to be updated.
Option 1: Manual update of test data
1. The instrument needs to be at the Main Menu.
2. Insert the blank or sample cuvette/vial (depending on the 
working procedure) in cell compartment #1 and close the cell 
compartment.
3. After "Barcode controlnumber?", a data overview is displayed, 
including the specifications of the test that is to be revised.
Note: The working procedures of the corresponding test contains the 
new data (Wavelength, Factors, Measurement Ranges, Conversion 
Factor for chemical Form, etc.).
4. Highlight the corresponding line and press Edit.
5. When finished programming, press Store and then Cancel to 
return to the User Programs menu.
6. Press Start to complete the test with the new data.
.
Option 2: Manual check/revision of test data
If an incorrect entry was made during the data update and is not 
recognized via the barcode of the cuvette/sample vial (e.g. an 
incorrect measuring range), another option is available to check 
and if necessary to update, the test data.
1. The instrument need to be at the Main Menu.
2. Insert the blank or sample cuvette/vial (depending on the 
working procedure) into the cell compartment #1 and close the 
cell compartment.
3. Press Options>More>Edit.
4. A data overview is displayed, including the specifications of the 
test that is to be revised. Compare the displayed data with the 
data in the working procedure and edit the displayed data in 
accordance with the working procedure.