Standard Operations
5.5.6.3 Program a new test
1. The instrument needs to be at the Main Menu.
2. Insert the sample cuvette/vial or blank (depending on the 
working procedure) into the cell compartment #1 and close the 
cell compartment.
3. After "Program not available", a data overview is displayed, 
including the specifications of the test that is to be revised.
The working procedures of the corresponding test contains the 
new data (Wavelength, Factors, Measurement Ranges, 
Conversion Factor for chemical Form etc.).
4. Highlight the corresponding line including the specifications of 
the test that is to be programmed and press Edit.
5. To finish programming, press Store and then Cancel to return 
to the Main Menu.
6. Press Start to run the test with the new data.
5.5.7 Upgrade of the instrument software
To obtain the software for the update from the Internet at 
www.hach.com:
1. Go to http://www.hach.com.
2. On the DR 5000 product page, click Lab System 
Software/Software Update Downloads under the heading 
Downloads.
3. Locate the appropriate download and follow the prompts for 
saving the file(s) to the USB memory stick.
4. On the DR 5000, press Instrument Update in the System 
Checks menu.
5. Connect the USB stick to the USB interface on the DR 5000 
(
). Press OK.The link is established 
automatically and the software is updated.