DB1 Liquid Level Measurement Sensor 
3:  Excitation with Delay (P22) 
 1:  1 
 EX 
Chan 
 
  2:  0 
  Delay w/EX (units=.01sec)  
  3:  150 
  Delay after EX (units=.01sec)  
 4:  0 
 mV 
Excitation 
 
 
Subroutine 1 measures the pressure in the tubing connected to the second value: 
 
4:  Do (P86) 
  1:  1 
  Call Subroutine 1  
 
5:  Do (P86) 
  1:  52 
  Set low Port 2  
 
The following instructions convert the mV readings to engineering units: 
 
6:  Z=X*Y (P36) 
  1:  8 
  X Loc AVG LEVEL 
 2:  4 
 Y 
Loc 
MULT. 
  3:  1 
  Z Loc [:LEVEL#1  ]  
 
7:  Z=X+Y (P33) 
  1:  1 
  X Loc LEVEL#1  
 2:  5 
 Y 
Loc 
OFFSET 
  3:  1 
  Z Loc [:LEVEL#1  ]  
 
8:  If time is (P92) 
  1:  0 
  minutes into a  
 2:  15 
 minute 
interval 
 
  3:  10 
  Set high Flag 0 (output)  
 
9:  Real Time (P77)  
 1:  110 
 Day,Hour-Minute 
 
 
10:  Average (P71) 
 1:  1 
 Rep 
 
 2:  1 
 Loc 
LEVEL#1 
 
 
11:  End Table 1 
 
* Table 3 Subroutines  
 
Pressure Transducer Measurement Routine: 
 
1:  Beginning of Subroutine (P85) 
 1:  1 
 Subroutine 
Number