Campbell Manufacturing LIQUID LEVEL DB1 用户手册

下载
页码 18
DB1 Liquid Level Measurement Sensor 
13:  Do (P86)  
  1:  1 
  Call Subroutine 1  
 
14:  Z=X (P31) 
  1:  8 
  X Loc AVG LEVEL  
  2:  9-- 
  Z Loc [:mV/V #1  ]  
 
15:  Do (P86) 
  1:  51-- 
  Set low Port 1  
 
16:  End (P95) 
 
17:  Z=X-Y (P35) 
  1:  10 
  X Loc mV/V #2  
  2:  9 
  Y Loc mV/V #1  
  3:  12 
  Z Loc [:DELTA V  ]  
 
Note:  On the next command, enter the measured Delta h value for F, using the units (e.g. inches, 
cm, etc.) you want the data to be in on your printout. 
 
18:  Z=F (P30)  
 1:  0 
 F 
 
  2:  0 
  Exponent of 10  
  3:  13 
  Z Loc [:DELTA H  ]  
 
19:  Z=X/Y (P38) 
  1:  13 
  X Loc DELTA H  
  2:  12 
  Y Loc DELTA V  
  3:  4 
  Z Loc [:MULT.    ]  
 
20:  Z=X*Y (P36) 
  1:  11 
  X Loc mV/V #3  
  2:  4 
  Y Loc MULT.  
  3:  5 
  Z Loc [:OFFSET   ]  
 
21:  Z=X*F (P37) 
  1:  5 
  X Loc OFFSET  
 2:  -1 
 F 
 
  3:  5 
  Z Loc [:OFFSET   ]  
 
22:  End (P95) 
 
23:  End Table 3 
 
6.  Maintenance 
The user should review the Installation and Maintenance section in the 
appropriate datalogger manual.  This section discusses maintenance required 
for components specific to the DB1 sensor. 
10