Technical Specifications
Unipolar Systems
High voltage Generator
Max. Output Voltage
160 kV
225 kV
Max. Output Current
45 mA
45 mA
Max. Output Power
4,5 kW, Limited by Tube Specification
4,5 kW, Limited by Tube Specification
High Voltage Ripple
5 V/mA (With High Voltage Cable 10 m), 40 kHz
5 V/mA (With High Voltage Cable 10 m), 40 kHz
Insulation
Oil
Oil
Housing Dimensions (Cathode) (W x D x H)
350 x 870 x 850 mm (13.8” x 34.3” x 33.5”)
350 x 870 x 850 mm (13.8” x 34.3” x 33.5”) 
Weight (Cathode)
189 kg (417 lbs), Including Power Module
189 kg (417 lbs), Including Power Module
tube voltage
Preselection and Setting
From 5 to 160 kV in 0.1 kV / 1 kV / 10 kV 
From 5 to 225 kV in 0.1 kV / 1 kV / 10 kV 
Digital Display of Set and Actual Values
Simultaneous 4 Digits Each
Simultaneous 4 Digits Each
Display Resolution
0.1 kV
0.1 kV
Accuracy
< ±1%
< ±1%
Reproducibility
±0.01% at Constant Temperature Level
±0.01% at Constant Temperature Level
Temperature Drift
< 65 ppm/°C
< 65 ppm/°C
tube Current
Preselection and Setting
From 0.1 to 45 mA in 0.01 mA / 0.1 mA / 1 mA 
From 0.1 to 45 mA in in 0.01 mA / 0.1 mA / 1 mA 
Digital Display of Set and Actual Values
Simultaneous 4 Digits Each
Simultaneous 4 Digits Each
Display Resolution
0.1 mA / 0.01 mA
0.1 mA / 0.01 mA
Accuracy
± 1%
±1%
Reproducibility
± 0.01% at Constant Temperature Level
± 0.01% at Constant Temperature Level
Temperature Drift
< 65 ppm/°C
< 65 ppm/°C
Exposure time
Programmable Timer
Non-Volatile Memory
Non-Volatile Memory
Preselection and Setting
From 0.1 to 99.9 Minutes in 0.1 Min. Increments or 
from 1 to 999 Sec. in 1 Sec. Increments or as direct 
Min./Sec. value (up to 99’59”)
From 0.1 to 99.9 Minutes in 0.1 Min. Increments or 
from 1 to 999 Sec. in 1 Sec. Increments or as direct 
Min./Sec. value (up to 99’59”)
Digital Display of Set and Actual Values
The Remaining Time Is Displayed, i.e. After a Mains 
Failure Exposure Can Be Continued Without any 
Time Error
The Remaining Time Is Displayed, i.e. After a Mains 
Failure Exposure Can Be Continued Without any 
Time Error
Prewarning time
Preselection and Setting
Digital Setting From 2 to 250 Seconds or 
de-activated
Digital Setting From 2 to 250 Seconds or 
de-activated
Programmed Mode
Number of Storable Programs
250
250
Warm-Up
Automatic Intelligent Tube conditioning based on 
Real-time Clock 
Extended warm-up for special conditioning
Automatic Intelligent Tube conditioning based on 
Real-time Clock 
Extended warm-up for special conditioning
X-ray Tube Setup
8 Tube selectable from a database of > 40 
pre-programmed tubes 
8 Tube selectable from a database of > 40 
pre-programmed tubes 
Operation History
512 Events (256 On/Off Events)
512 Events (256 On/Off Events)
Warm-up History
128 Events
128 Events
Control Module
Dimensions (W x D x H)
460 x 270 x 100 mm (18.1” x 10.6” x 3.9”)  
Built into Desk Housing
460 x 270 x 100 mm (18.1” x 10.6” x 3.9”) Built into 
Desk Housing
Weight
4.9 kg (10.8 lbs) Including Desk Housing
4.9 kg (10.8 lbs) Including Desk Housing
Connected loads
Power connection
1N PE 230 V ± 10% 50/60 Hz  
3N PE 400/230 V ±10%, 50/60 Hz, 3-Phase, 
Grounded Neutral TN-S or TN-C-S Mains (Star 
Connected System, Optional 3-Phase Isolation 
Transformer)
1N PE 230 V ± 10% 50/60 Hz  
3N PE 400/230 V ±10%, 50/60 Hz, 3-Phase, 
Grounded Neutral TN-S or TN-C-S Mains (Star 
Connected System, Optional 3-Phase Isolation 
Transformer)
Grounding
Separate Grounding for X-ray Tube and High 
Voltage Generator (Minimum 6 mm
2
)
Separate Grounding for X-ray Tube and High 
Voltage Generator (Minimum 6 mm
2
)
Mains Fuses
63 A (1N PE) or 25 A (3N PE) Time-Delay Fuses, 
Customer-Supplied
63 A (1N PE) or 25 A (3N PE) Time-Delay Fuses, 
Customer-Supplied
Operating Temperature Range
0°C to +40°C
0°C to +40°C
Storage Temperature Range
-30°C to +70°C
-30°C to +70°C
Bipolar Systems
High voltage Generator
Max. Output Voltage
-160 kV (Cathode), +160 kV (Anode)
-225 kV (Cathode), +225 kV (Anode)
Max. Output Current
45 mA
45 mA
Max. Output Power
4,5 kW (Cathode)
3 kW (Anode)
Limited by Tube Specification
4,5 kW (Cathode)
3 kW (Anode)
Limited by Tube Specification
High Voltage Ripple
10 V/mA (With High Voltage Cable 10 m), 40 kHz
10 V/mA (With High Voltage Cable 10 m), 40 kHz
Insulation
Oil
Oil
Operation History
512 Events (256 On / Off events)
512 Events (256 On / Off events)
Warm-up History
128 Events
128 Events
Housing Dimensions (Cathode) (W x D x H)
350 x 870 x 620 mm (13.8” x 34.3” x 24.4”)
350 x 870 x 620 mm (13.8” x 34.3” x 24.4”)
350 x 870 x 850 mm (13.8” x 34.3” x 33.5”)
350 x 870 x 850 mm (13.8” x 34.3” x 33.5”)
Weight (Anode)
123 kg (272 lbs)
123 kg (272 lbs)
Weight (Cathode)
189 kg (417 lbs), Including Power Module
189 kg (417 lbs), Including Power Module
tube voltage (Anode)
Preselection and Setting
From 5 to 320 kV in 0.1 kV / 1 kV / 10 kV 
From 5 to 450 kV in 0.1 kV / 1 kV / 10 kV 
Digital Display of Set and Actual Values
Simultaneous 4 Digits Each
Simultaneous 4 Digits Each
Display Resolution
0.1 kV
0.1 kV
Accuracy
< ±1%
< ±1%
Reproducibility
±0.01% at Constant Temperature Level
±0.01% at Constant Temperature Level
Temperature Drift
< 65 ppm/°C
< 65 ppm/°C
tube Current
Preselection and Setting
From 0.1 to 45 mA in 0.01 mA / 0.1 mA / 1 mA 
From 0.1 to 45 mA in 0.01 mA / 0.1 mA / 1 mA 
Digital Display of Set and Actual Values
Simultaneous 4 Digits Each
Simultaneous 4 Digits Each
Display Resolution
0.1 mA / 0.01 mA
0.1 mA / 0.01 mA
Accuracy
± 1%
±1%
Reproducibility
± 0.01% at Constant Temperature Level
± 0.01% at Constant Temperature Level
Temperature Drift
< 65 ppm/°C
< 65 ppm/°C
Exposure time
Programmable Timer
Non-Volatile Memory
Non-Volatile Memory
Preselection and Setting
From 0.1 to 99.9 Minutes in 0.1 Min. Increments or 
From 1 to 999 Sec. in 1 Sec. Increments or as 
direct Min./Sec. value (up to 99’59”)
From 0.1 to 99.9 Minutes in 0.1 Min. Increments or 
From 1 to 999 Sec. in 1 Sec. Increments or as 
direct Min./Sec. value (up to 99’59”)
Digital Display of Set and Actual Values
The Remaining Time Is Displayed, i.e. After a Mains 
Failure Exposure Can Be Continued Without any 
Time Error
The Remaining Time Is Displayed, i.e. After a Mains 
Failure Exposure Can Be Continued Without any 
Time Error
Prewarning time
Preselection and Setting
Digital Setting From 2 to 250 Seconds or 
de-activated
Digital Setting From 2 to 250 Seconds or 
de-activated
Programmed Mode
Number of Storable Programs
250
250
Warm-Up
Automatic Intelligent Tube conditioning based on 
Real-time Clock 
Extended warm-up for special conditioning
Automatic Intelligent Tube conditioning based on 
Real-time Clock 
Extended warm-up for special conditioning
X-ray Tube Setup
8 Tube selectable from a database of > 40 
pre-programmed tubes 
8 Tube selectable from a database of > 40 
pre-programmed tubes 
Control Module
Dimensions (W x D x H)
460 x 270 x 100 mm (18.1” x 10.6” x 3.9”)  
Built into Desk Housing
460 x 270 x 100 mm (18.1” x 10.6” x 3.9”)  
Built into Desk Housing
Weight
4.9 kg (10.8 lbs) Including Desk Housing
4.9 kg (10.8 lbs) Including Desk Housing
Connected loads
Power connection
1N PE 230 V ± 10% 50/60 Hz  
3N PE 400/230 V ±10%, 50/60 Hz, 3-Phase, 
Grounded Neutral TN-S or TN-C-S Mains (Star 
Connected System, Optional 3-Phase Isolation 
Transformer)
1N PE 230 V ± 10% 50/60 Hz  
3N PE 400/230 V ±10%, 50/60 Hz, 3-Phase, 
Grounded Neutral TN-S or TN-C-S Mains (Star 
Connected System, Optional 3-Phase Isolation 
Transformer)
Grounding
Separate Grounding for X-ray Tube and High 
Voltage Generator (Minimum 6 mm
2
)
Separate Grounding for X-ray Tube and High 
Voltage Generator (Minimum 6 mm
2
)
Mains Fuses
63 A (1N PE) or 25 A (3N PE) Time-Delay Fuses, 
Customer-Supplied
63 A (1N PE) or 25 A (3N PE) Time-Delay Fuses, 
Customer-Supplied
Operating Temperature Range
0°C to +40°C
0°C to +40°C
Storage Temperature Range
-30°C to +70°C
-30°C to +70°C
11