4.5.1.4  Diagnostic Error Messages 
Recorders with displays run background tests as described in Section 8 – Troubleshooting and Pen 
Alignment of Recorders with Display 
to verify data and memory integrity. If there is a malfunction, an error 
message will blink in the lower display. If more than one self-diagnostic test fails, the display will cycle 
through the appropriate error messages.    
A list of background test error messages is in Table 4-28.  
If any of these error messages are displayed, refer to Section 8 for information on correcting the problem. 
In addition, an error message will be displayed if the operator tries to run a setpoint program in which the 
number of the ending segment is lower than the segment number of the start segment.  See Appendix C for 
details. 
 
Table 4-28  Error Messages 
Prompt Description 
CALERR
 
Indicates failure of the calibration test. 
CFGERR 
Indicates failure of the configuration test. 
EEFAIL 
Unable to write to nonvolatile memory 
FAILSF 
Failsafe value being used as output 
INP FL 
Two consecutive failures of input integration 
INPRNG 
Input out-of-range 
P1 ERR 
or 
P2 ERR 
Indicates error on channel other than the one on display. 
RAMTST 
Indicates failure of the RAM test. 
SEGERR 
Indicates that the number of the setpoint program’s end segment is lower 
than the segment number of the start segment. 
 
ATTENTION 
In addition to error messages, the recorder will display special status messages when optional 
features are active. 
•  If the lower display flashes “TUNING”, the recorder is in the process of automatically 
determining tuning parameters using the Accutune II feature described in Appendix D.   
•  If the lower display flashes “RUN n” or “HOLD”, the recorder is executing setpoint program “n” 
as described in Appendix C, and the run or hold action was initiated locally. 
•  If the lower display flashes “run n” or “hold”, the recorder is executing setpoint program “n”, 
and the run or hold action was initiated remotely.