FP-RTD-124 and cFP-RTD-124
Figure 3 shows the input circuitry on one channel.
Figure 3.  FP-RTD-124 Input Circuit
Taking Measurements from Four-Wire RTDs
Four-wire RTDs generally have two wires of one color and two of 
another color (usually red and white). Connect the wires of one 
color to the positive terminals (EXCITE+ and SENSE+). Connect 
the wires of the other color to the negative terminals (COM and 
SENSE–).
Figure 4.  Four-Wire RTD Connections on One Channel
Taking Measurements from Three-Wire RTDs
NI recommends the [c]FP-RTD-122 for three-wire RTD 
measurements because it includes special sensing circuitry to 
compensate for up to 95% of the lead resistance of the third wire. 
Most three-wire RTDs have two wires of one color (usually red, 
sometimes black) and one wire of another color (usually white, 
sometimes red). Connect the two wires of the same color to the 
SENSE+
[c]FP-RTD-124
SENSE–
COM
EXCITE+
Amplifier
Pulsed 2 mA
Four-Wire
RTD
16-bit
ADC
SENSE+
[c]FP-RTD-124
SENSE–
COM
EXCITE+
Four-Wire
RTD