National Instruments NI 5102 用户手册

下载
页码 87
Chapter 3
Digitizer Basics
© National Instruments Corporation
3-9
Analytically, obtaining a flat frequency response means:
R
in
/(R
in
+ R
p
) = Cp/(C
p
+ C
in
+ C
c
)
It can be shown that:
R
in
(C
in
+ C
c
) = C
p
R
p
or the time constant of the probe equals the time constant of the digitizer
input.
How to Compensate Your Probe
Adjusting the tunable probe capacitor to get a flat frequency response
is called probe compensation. On the NI 5102, you can select a 0–5 V,
1 kHz pulse train as reference to output on PFI1 or PFI2. Calling
niScope_ProbeCompensation
in LabVIEW, however, will only allow
output to PFI1. Refer to Figure 3-7 as you follow these instructions to
compensate your probe:
1.
Connect the BNC end of the probe to an input channel, either CH0
or CH1 and select X10 attenuation on the body of the probe tip.
2.
Attach the BNC adapter (probe accessory) to the tip of the probe.
3.
Connect the SMB100 probe-compensation cable to one of the
PFI lines. On the NI 5102 (PXI) this line is PFI1.
4.
Attach the probe with the BNC adapter to the BNC female end of the
SMB100 cable.
5.
Enable the probe compensation signal on the PFI line you selected in
step 3. See your application software documentation for more
information how to perform this step.
6.
Digitize data on the input channel, amplifying the signal until the
signal starts to clip. Then go back one step so it does not clip anymore.
This step ensures that you use the main dynamic range of the ADC.
7.
Adjust the tunable capacitor to make the waveform look as square as
possible.
8.
For the most accurate measurements, compensate probes for each
channel (CH0 and CH1) and use them on that channel only.
Recompensate when using the same probe on a different channel.