Chapter 3
Digitizer Basics
© National Instruments Corporation
3-11
Figure 3-8. Probe Compensation Comparison
Active and Current Probes
You can also use active probes and current probes with digitizers
and DSOs.
Active probes, such as differential and field-effect transistor (FET) probes,
contain active circuitry in the probe itself to reject noise and amplify the
signal. FET probes are useful for low-voltage measurements at high
frequencies and differential probes are noted for their high common mode
rejection ration (CMRR) and nongrounded reference.
Instead of using a series resistance in the loop to measure current, current
probes magnetically measure AC and/or DC current flowing in a conductor.
This lack of series resistance causes very little interference in the circuit
being tested.
Probe Adjustment Signal
Probe Adjustment Signal
Probe Adjustment Signal
Proper Amplitude of a
1 MHz Test Signal
Reduced Amplitude of a
1 MHz Test Signal
Increased Amplitude of a
1 MHz Test Signal
a. Compensated Correctly
b. Undercompensated
c. Overcompensated