National Instruments NI 5102 用户手册

下载
页码 87
Chapter 4
Hardware Overview
4-10
ni.com
Table 4-5 describes the posttrigger acquisition signals.
Table 4-5. Posttrigger Acquisition Signals
Signal
Description
Start Trigger
Triggers the acquisition. It can be generated through software, or CH0,
CH1, TRIG, PFI1, and PFI2, or any of the seven RTSI bus trigger lines.
RTSI bus trigger lines are available only on the NI 5102 (PCI, PXI, ISA).
Scan Clock
Causes the ADC to convert the input signal into digital data. This signal is
also used in the memory controller to write the data into onboard memory.
This signal can be generated internally, with a 24-bit counter clocked with
a 20 MHz signal to generate pulses from 20 MHz to 1.19 Hz. The 24-bit
counter provides a wide choice of valid frequencies for the Scan Clock
signal. In addition, Scan Clock can also be selected from CH0, CH1, TRIG,
PFI1, and PFI2, or any of the seven RTSI bus trigger lines. RTSI bus trigger
lines are available only on the NI 5102 (PCI, PXI, ISA).
End of Acquisition
Indicates end of acquisition to the control logic in the hardware. It is
generated from a counter that keeps track of the number of points remaining
in the acquisition. It can be exported from the device on the PFI lines.