Chapter 4
Hardware Overview
© National Instruments Corporation
4-11
Pretrigger Acquisition
In pretrigger acquisition mode, the device acquires a certain number of
scans, called the pretrigger scan count, before the trigger occurs. After
satisfying the pretrigger scan count requirement, hardware keeps acquiring
data and stores it in a circular buffer implemented in onboard memory. The
size of the circular buffer equals the pretrigger scan count. When the trigger
occurs, hardware acquires and stores the posttrigger scans and the
acquisition terminates. Table 4-6 shows the minimum and maximum
number of samples available on the NI 5102 in pretriggered mode.
Note
If Scan Clock is externally supplied, a free-running clock must be used for proper
operation.
Table 4-6. Possible Number of Samples for Pretriggered Mode
Number of Channels
NI 5102 (PCI, PXI)
NI 5102 (ISA, PCMCIA, USB)
Min
Max
Min
Max
One
Pretriggered scans
Posttriggered scans
1
1
663,000
16,777,216*
1
1
663,000 – A
663,000 – B
Two
Pretriggered scans
Posttriggered scans
1
1
331,500
16,777,216*
1
1
331,500 – A
331,500 – B
* Dependent on available memory
A
The number of posttriggered scans
B
The number of pretriggered scans