RayTek 54301 用户手册

下载
页码 93
Programming Guide 
11.4 Device Setup 
11.4.1 Temperature Calculation 
U=C 
unit for the temperature value 
E=0.950 
Emissivity  setting  (Caution:  according  to  the  settings 
for  “ES”,  see  section  11.4.2 
 on page 69.) 
XG=1.000 
Setting for transmission 
For  the  calculation  of  the  temperature  value,  it  is  possible  to  set  an 
offset (relative number to be added to the temperature value), and a 
gain value. 
DG=1.0000 
Gain adjustment for the temperature signal 
DO=0 
Offset adjustment for the temperature signal 
In case the ambient temperature is not requested by the internal head 
temperature,  you  must  set  the  ambient  temperature  values  as 
follows: 
A=250.0 
Ambient temperature (example) 
AC=1 
Control ambient background temp. compensation 
11.4.2 Emissivity Setting and Alarm Set points 
The device allows three choices for the emissivity setting and two for 
the alarm output setting. 
ES 
Selection of the emissivity setting. 
ES=1 
Emissivity  set  by  a  constant  number  according  to  the 
„E“ command 
ES=E 
Emissivity set by a voltage on FTC1 (analog input) 
ES=D 
Emissivity  set  by  the  entries  in  a  table  (selected  by 
digital inputs FTC1 – FTC 3) 
?CE 
asks  for  the  emissivity  value  that  is  actually  used  for 
temperature calculation 
MID 
69